Wire 2/Pair
11 (CB): Red
RS-232 RX
Y
NC
B
Triad Black
Wire
Z
Wire 2/Pair 9
(CB): Gray
RS-232 GND
Z
NC
C
Triad White
Wire
a
1 M to
Power
Bundle
Shield Drain
Wire
Power Bundle
Shield/CASE Ground
a
Power Bundle Shield
Drain Wire
b
Wire 1/Pair
12 (PB):
Black
Input Power +
(9-36V)
b
Wire 1/Pair 12 (PB):
Black
c
Wire 2/Pair
12 (PB):
Orange
Input Power Return
c
Wire 2/Pair 12 (PB):
Orange
D
Triad Shield
Drain Wire
E
NC
F
NC
Table 2: Connection details of standard 6 meter cable.
Operational Control of the EpiSensor2
The sensor can be controlled in multiple ways: button pushing (local), digitizer enable lines (remote) and
RS-232 (local and/or remote). Using either of these, the operator can control the full-scale range, and
AUTOZERO state, of the sensor. Using the digitizer or RS-232 interfaces, the user can exercise a
complete calibration (CAL) capability. Using a dedicated digital line connected to the digitizer, the
operator can issue a RESET to the sensor that is similar to a Power-ON-RESET. Finally, the user can use
the RS-232 interface to communicate with the sensor to receive a variety of sensor-specific information
(serial numbers, scalar response values, etc.). The RS-232 command line menu is discussed further
below.