background image

 

 

 

 

KILEWS

 

 

 
                     

                                                                                  KILEWS    JAPAN    CO., LTD. 

 

 

                                                         

 KTM - CSH 

                                         

 

 

 

 

 

 

 

 

数字扭力计操作手册

 

DIGITAL TORQUE METER INSTRUCTION MANUAL 
                                                                                                                           

        OSH-Q04 

 

Содержание KTM-CSH

Страница 1: ...KILEWS KILEWS JAPAN CO LTD KTM CSH DIGITAL TORQUE METER INSTRUCTION MANUAL OSH Q04...

Страница 2: ...Digital Torque Meter KILEWS 1 2 24 3 4 5 6 7 8 9 10 11 8 12 13 14 2...

Страница 3: ...Digital Torque Meter KILEWS 15 16 17 18 19 3...

Страница 4: ...G LOBAT low battery lamp NG LOBAT DOWN Switch Lamp GOOD Torque detection part ON OFF Key Measurement Mode change switch AC DC Adaptor Terminal SELECT Key CLEAR Key Measurement Unit change switch RS232...

Страница 5: ...KTM CSH 30 174 Lbf in 30 200 Kgf cm 3 20 N m 200 1000 0 5 15 199 1 LCD P P PEAK TRACK P D PEAK DOWN RS232 9600BPS 400 6VDC Ni Cd 1 2V x5 Lbf in Kgf cm N m 6 12 300 AC 110V or 220V DC 7 25V 120mA 180...

Страница 6: ...3 5 4 ON OFF LOBAT NG ON OFF 1 1 TRACK 2 0 0 3 PD TR PP 10 5 5 1 SELECT UP UP DOWN N G 2 SELECT UP DOWN GOOD N G N G GOOD 3 SELECT UP DOWN 15 4 SELECT AC UP DOWN 0 1 3 0 0 5 0 0C 0 1C 0 5C 1 0C 1 5 2...

Страница 7: ...tal Torque Meter KILEWS 5 6 1 2 ON OFF LOBAT NG 3 4 5 6 7 CLEAR RS232C PD PP 8 5 ON OFF 1 10 LOBAT 6 PEAK PEAK 7 1 P P P D RS 232C RS 232C 7 2 7 3 1 SELECT CLEAR FA DOWN UP 2 SELECT LA 3 SELECT P 4 Cl...

Страница 8: ...Digital Torque Meter KILEWS 8...

Страница 9: ...e battery be sure to use only exclusive charger Using a charger other than specified may cause a fire or injury 11 Charge properly Charge on the indicated voltage Do not use DC power or engine generat...

Страница 10: ...it may not only show the full performance but also may cause any accident or injury 18 Use only the specified accessories and attachments Do not use any accessories and attachments other than ones spe...

Страница 11: ...orque Display Part UP Switch NG LOBAT low battery lamp DOWN Switch Lamp Torque detection part ON OFF Key Measurement Mode change switch AC DC Adaptor Terminal SELECT Key CLEAR Key Measurement Unit cha...

Страница 12: ...h Result of calibration document Certification on calibration document Traceability system figure 4 Explanation of the sign displayed on LCD Name Representation high value The high value of the succes...

Страница 13: ...of setting value It is performed a additional tightening It may be unable to measure depending on a situation 5 4 ON OFF of Power Supply Turn on a power supply switch ON OFF Supposing a LOBAT NG lamp...

Страница 14: ...he zero in manual operation it becomes clear 5 End of a setup If the SELECT key is pressed again it is displayed as and is a setting end Green LED GOOD lamp puts out the light The inside of measuremen...

Страница 15: ...er torque LED red Blink green lighting red Blink red lighting Detection check peak down LED green lighting 7 Memory function and a data output 7 1 Usage of Memory Data is always saved whenever it pres...

Страница 16: ...emory memorizes it outputs a signal also to a RS232C output and a printer output 7 5 Data Format The data output is RS 232C output conformity Data can be outputted to the printer corresponding to RS 2...

Страница 17: ...formed please perform zero adjustment P7 5 6 Moreover since all the contents of a memory are eliminated please redo a setup once again 8 2 Power Supply If the LOBAT NG lamp of a display machine blinks...

Страница 18: ...not turned on Is charge carried out Please carry out regular time charge using the exclusive charger attached Even if it turns on a power switch a LOBAT lamp blinks and a power supply is shut off imme...

Страница 19: ...Digital Torque Meter KILEWS 19...

Отзывы: