TOS9200/9201
Basic Operations 3-67
Setting the program name
Enter the name of the program to be created (using up to 12 characters).
1.
Press the SHIFT + F1 keys to move the cursor to the program name.
2.
Using the keys, move the cursor to the target character.
3.
Using the rotary knob, select characters. Use ASCII characters from
20H (space) to 7EH (~) (see the Appendix 2).
4.
Upon completion of editing, press the SHIFT + F1 keys to move the cur-
sor to the step number.
Setting the test conditions (Memory No.) and interval time
• If a test condition under the preset memory number is set to TIMER OFF, a pro-
gram test ends once the STOP switch is pressed, preventing the next test from
being performed.
For each step, set the memory number of the test conditions to be used and the inter-
val between steps. The program executes the test in order of step number.
When the cursor is positioned at the step number, the screen can be scrolled using
the rotary knob. The keys can be used to perform scrolling, regardless of the
location of the cursor.
Upon completion of each step, “END” or “RET” is displayed.
1.
To move the cursor to the step number, press the SHIFT + F1 keys.
2.
Move the cursor to the step below the step to be inserted. To add a step
to the previous step, move the cursor to END or RET.
3.
Press the F1 key (INS). A step (ACW00-0.2s) will be inserted at the cur-
sor.
4.
Using the key, move the cursor to ACW to the right of the step num-
ber.
5.
Using the rotary knob, set the target test (ACW, DCW, IR, EC). (To con-
trol the earth continuity tester TOS6200, select EC. See "Chapter 5
6.
Using the key, move the cursor to the memory number.
7.
Using the rotary knob, set a memory number.
8.
Using the key, move the cursor to the interval time to the right of the
memory number.
9.
Using the rotary knob, set an interval time (from 0.2 s to 9.9 s, HOLD).
If the interval time is set to HOLD, the next step starts when the START switch
is pressed with the speci
fi
ed step in the HOLD status.
To delete a step, move the cursor to the corresponding step number and press the F2
key (DEL).
NOTE
Содержание TOS9200
Страница 22: ...P 10 Preface TOS9200 9201...
Страница 116: ...3 78 Basic Operations TOS9200 9201...
Страница 130: ...4 14 Using Terminals and Connectors TOS9200 9201...
Страница 148: ...7 4 Maintenance TOS9200 9201...
Страница 162: ...8 14 Specifications TOS9200 9201...
Страница 170: ...I 4 Index TOS9200 9201...
Страница 171: ......