8
WARNINGS AND CAUTIONS FOR OPERATING THE TESTER
TOS8870A
2.2.5 Suspension of testing
Except under testing, the TEST VOLTAGE dial shall be turned to the extreme left ("0" position).
Also, push the STOP switch (H.V OFF) to ensure safety.
The Power switch shall be turned off (
○
) if the Tester is not used for some time or the operator is to
leave from the Tester.
Fig.2-3
2.2.6 Critical areas of the Tester under operation
It is dangerous to touch such high voltage areas as the DUT, and the test lead-wire, probe and output
terminal while the Tester is under operation.
Fig.2-4
•
NEVER touch the alligator clip of the test lead-wire and the vinyl-cover of the clip
because they are NOT insulation proof.
2.3 Warning after the output has been cut off
2.3.1 Con
fi
rmation on completion of testing
You may touch the DUT and the high-voltage areas (test lead-wire, probe or output terminal) for
correction of the wiring or any other purpose provided that the following con
fi
rmation has been
made;
- the indication of output voltage is at "0", and
- the DANGER lamp has been turned off.
In the case of insulation resistance test, DUT is charged after testing.
Be sure to read "2.3.2 Cautions in electric charging in insulation resistance test".
AUTO
MANUAL
IR
ACW
ACW
IR
ACW
IR
T IMER
OFF
ON
ADD
LO WER
UPPER
PE
S
50
1
2
2
-2
UPPER
LOWER
OUTPUT
INSULATION
RESISTANCE
WITHSTANDING VOLTAGE
REMOTE
ACW
FAIL
AUTO TEST
PROTECT ION
PASS
IR
FAIL
500V
1000V
2.5kV
5kV
FUNCTION
REFERENCE
OFF
ON
CUTOFF CURRENT (mA)
LOW
5kV
1kV
MA X
HIGH VOLTAGE
MAX
0
TEST VOLTAGE
AC WI THSTANDING VOLTAGE INSUL ATI ON RESISTANCE TESTER
TO S 8870 A
S
TIMER
S
TIMER
/
PE
S50122-1
1000V
RANGE
500V
REFERENCE
OFF
ON
REFERENCE
8
0.5
1
2
4
10
100
5kV
RANGE
2.5kV
x 0.1
x1
x1
x 0.1
PE
S5
01
22
-4
POWER
DANGER
9 9 9
9 9 9
TEST VOLTAGE dial
STOP switch
(H.V OFF)
Power switch
WARNING
Содержание TOS8870A
Страница 53: ...TOS8870A SPECIFICATIONS 41 7 4 Outline Drawing Unit mm Fig 7 1...
Страница 56: ...TOS8870A OPERATION MANUAL...