46 Specifications
TOS8030
Chapter 9
Specifications
Provides the electrical and mechanical specifications for the TOS8030.
The specifications are based on the following conditions and settings, unless other-
wise specified.
•
Warm-up time: 30 minutes
•
Temperature: 5
°
C to 35
°
C
•
Relative humidity: 20% to 80% (with no dew condensation)
•
“xx% of reading” represents xx% of voltmeter reading.
9.1
Basic Specifications
*1 Time limitations on the output
The heat radiation capacity of the output voltage generator section of the tester is designed to be 1/2
of the rated output, in consideration of the instrument dimensions, weight, costs, and other factors.
The tester, therefore, must be used under the following time constraints (interval time and output
time). If used beyond these limits, the output section may overheat, activating the internal protection
circuit. In such cases, always halt testing for a duration equal to or greater than the test duration.
*2 Test voltage waveform
If AC voltage is applied to a capacitive load, the output voltage in certain cases may rise above the value at
no-load, depending on the value of the capacitive element of the load. Moreover, for samples whose
capacitance values show voltage dependency (as with ceramic capacitors), waveform distortions may
result. However, for a test voltage of 1.5 kV, the effects of a capacitance of 1000 pF or less may be ignored.
Output section
Output voltage range
0.05 kV to 3.00 kV/single range
Maximum rated load (*1)
30 VA (3 kV/10 mA) (at a nominal input rating)
Output voltage waveform (*2)
AC line waveform
Voltage regulation
20% or less (during transition from the maximum rated load to no-load)
Switching
A zero-start switch is used.
Voltmeter
Display
Digital three-digit indication
Measurement range
0.00 kV to 4.00 kV
Display resolution
10 V
Accuracy
± 1.5% FS or
Vm
≥
1.00 kV: ± (5% of reading),
Vm < 1.00 kV: ± (5% of r 30 V)
– whichever is smaller.
where FS: full scale (4.00 kV), Vm: measured voltage value
Response
Mean value response/rms value indication
Ambient
temperature (t)
Reference value (i)
Interval time
Output time
t
≤
40
°
C
5 mA < i
≤
11 mA
Equal to or greater than
output time
10 minutes max.
i
≤
5 mA
Not necessary
Continuous output is possible.