TOS7200
33
System Setting
In addition to test conditions such as the test voltage, this product has settings related to the
system.
In system setting, the following items can be set:
•
Double action
•
Momentary
•
Fail mode
•
Pass hold
•
Average count (AVG)
•
Auto-range
•
Lower judgment (LOWER ON)
•
Communication rate (SPEED)
•
Buzzer volume (BUZZER)
Controlling the system settings
1
Press the STOP switch.
2
Check that the DANGER lamp is turned off.
3
With the SHIFT key held down, press the SYSTEM key.
The TOS7200 enters system setting mode.
4
Press the or key to move the cursor to the digit at which a value is to
be set.
The 7-segment LED corresponding to the system setting function display turns
on to indicate that a system setting item has been selected.
5
With the SHIFT key held down, press the or key to change the value.
To exit system setting
1
Press the STOP switch.
The display returns to the previous status before the tester entered system
setting.
MEMORY
VOLTAGE
RESISTANCE
TIMER
MEMORY
VOLTAGE
RESISTANCE
TIMER
System setting display
SYSTEM key
For moving the
cursor or changing
a numerical value