4-12
Basic Operation
TOS6200A/6210
4.2.6
Offset Canceling Function
The offset canceling function displays the measured value after subtracting a pre-
stored offset value from it. This function is useful for cases where a surplus resis-
tance component (such as a test-lead resistance component or the OUTPUT termi-
nal’s contact resistance) is to be canceled from the measured value in two-terminal
measurements.
To use the offset canceling function, you need to measure the surplus resistance
component (offset value) in advance using this tester.
Offset Canceling Function ON/OFF Settings
To use the offset canceling function, set the OFFSET indication to ON.
Pressing the SHIFT and F3 keys together allows you to toggle the offset canceling
function ON/OFF. Pressing the SHIFT + F3 toggles between ON and OFF.
You can also use the
▲▼
keys to move the cursor to the offset function ON/
OFF indicator. With the cursor at the OFFSET indication use the rotary knob to set.
Turning the knob clockwise:
ON
Turning the knob counterclockwise: OFF
25.0
A
0.100 60.0
s
MAIN
MAIN
FREQ
LOWER
OFF
OFFSET
ON
TIMER
ON
READY
50Hz
F1 F2 F3 F4 F5
SHIFT
FREQ
50/60Hz
LOWER
ON/OFF
OFFSET
ON/OFF
TIMER
ON/OFF
UPPER
JUDGE
R/V
▲
▲
Содержание TOS6200A
Страница 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Страница 14: ...1 8 General TOS6200A 6210...
Страница 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Страница 34: ...3 8 Part Names and Functions TOS6200A 6210...
Страница 76: ...4 42 Basic Operation TOS6200A 6210...
Страница 134: ...7 6 Maintenance TOS6200A 6210...
Страница 150: ...A 8 Appendix TOS6200A 6210...
Страница 153: ...TOS6200A 6210 Index I 3 U Unpacking 2 2 UP LOW 4 35 Upper reference value 4 5 V VOLT LIMIT 4 39...