TOS6200A/6210
Installation and Preparation for Use
2-5
2.4
Connecting the Test Leads
• This product carries a maximum current flow of 30 A or 62A. Always check
to make sure that no connections are loose. Loose connection will result in
overheating of the OUTPUT terminals or the DUT (Device Under Test),
which may then result in burns or injury.
• Never connect the voltage measurement cable (thin wire) of the supplied
test leads or optional test probe to the OUTPUT terminals. The nominal
sectional area of this wire is inadequate for such currents, and burning
may result.
• This product carries a large current and consequently generates a strong
magnetic field. Make sure that no articles that may be affected by mag-
netic fields are located near the test leads or current output lines.
For example, images on a CRT positioned close by may be significantly
distorted.
2.4.1
Supplied Test Lead TL11-TOS/TL12-TOS
1.
Remove the short-circuit bars connecting the OUTPUT and SAMPLING
terminals.
2.
Connect the current output line on the crimp terminal side of the test
lead (black or white) to the LOW side of the OUTPUT terminals; con-
nect the voltage measurement line on the banana plug side to the LOW
side of the SAMPLING terminals.
Check that the connections are
secure.
WARNING
CAUTION
TOS6200A
Take care not to get your finger
caught between the output
terminals when turning the knobs.
TOS6210
Содержание TOS6200A
Страница 1: ...Part No IB028661 Mar 2015 USER S MANUAL EARTH CONTINUITY TESTER TOS6200A TOS6210...
Страница 14: ...1 8 General TOS6200A 6210...
Страница 26: ...2 12 Installation and Preparation for Use TOS6200A 6210...
Страница 34: ...3 8 Part Names and Functions TOS6200A 6210...
Страница 76: ...4 42 Basic Operation TOS6200A 6210...
Страница 134: ...7 6 Maintenance TOS6200A 6210...
Страница 150: ...A 8 Appendix TOS6200A 6210...
Страница 153: ...TOS6200A 6210 Index I 3 U Unpacking 2 2 UP LOW 4 35 Upper reference value 4 5 V VOLT LIMIT 4 39...