TOS5300_INTERFACE
Tutorial
Performing tests
We will use an AC withstanding voltage test (ACW) to explain how to perform single operations.
Set the test conditions (test voltage, limit voltage, upper and lower limits, and test time):
SOUR:VOLT 1.5KV
SOUR:VOLT:PROT 2KV
SENS:JUDG 10MA
SENS:JUDG:LOW 0.01MA
SENS:JUDG:LOW:STAT ON
SOUR:VOLT:TIM 60S
SOUR:VOLT:TIM:STAT ON
Set additional test conditions (start voltage, voltage rise time, voltage fall time, and test voltage
frequency):
SOUR:VOLT:STAR:STAT ON
SOUR:VOLT:SWE:TIM 5S
SOUR:VOLT:SWE:FALL:TIM:STAT OFF
SOUR:VOLT:FREQ 60HZ
After setting the test conditions, start the test.
First, set the mode to perform the test in:
SOUR:FUNC:MODE ACW
Start the test (in this example, we will just start the test without using any triggers).
When you send the INIT command, the test starts.
You cannot start the test if memory is being recalled or saved, the TOS5300 Series is in protection
mode, or PASS/FAIL judgment is being performed.
TRIG:TEST:SOUR IMM
TEST:EXEC
You can query the voltage and current values during testing:
MEAS:CURR?
MEAS:VOLT?
Testing stops when the test time elapses (a PASS judgment), if a FAIL judgment occurs, or if the
STOP switch is pressed. You can also end testing by sending the ABOR command, but the test
results will be discarded. When testing finishes, you can query the test results (the overall
judgment).
The condition of the OPER:TEST register is returned in NR1 format:
STAT:OPER:TEST:COND?