20
TOS5300_INTERFACE
Test Mode Settings
*TRG
IEEE 488.2-1992, section
10.37
Trigger command.
This is a substitute command for the IEEE 488.1 get message (Group Execute Trigger). If the
TOS5300 Series is in a state in which it does not accept triggers, an SCPI error (-211, “Trigger
ignored”) occurs.
Command
*TRG
*TST
IEEE 488.2-1992, section
10.38
Executes a self-test. You can query which error occurred by sending the SYST:ERR? command.
Command
*TST?
Response
Returns “0” if no errors are detected. Returns the error code if an error is detected.
*WAI
IEEE 488.2-1992, section
10.39
Prevents the TOS5300 Series from executing subsequent commands until all operations that are in
standby have been completed.
Command
*WAI
Test Mode Settings
SOUR:FUNC:MODE
Sets the test mode. You can only set the test mode to one of the test modes that is available on your
model.
Command
SOURce:FUNCtion:MODE {ACW|DCW|IR}
SOURce:FUNCtion:MODE?
Parameter
Value:
ACW
AC withstanding voltage test (default)
DCW
DC withstanding voltage test
IR
Insulation resistance test
Response
Returns the test mode in <character> format.
See
See
See