16
TOS5300_INTERFACE
Command Description in This Manual
In this manual, commands are described in the following manner.
References to command descriptions
Append the value that you want to set the
setting to after the command.
To set the test voltage of a DC withstanding
voltage test (DCW) to 6.2 kV, send
SOUR:DCW:VOLT 6.2k.
Commands that have these marks are affected
when an *RST or *RCL command is sent. The
settings for the command are changed to the values
that are shown in the “Default values” section.
The test voltage is changed to 0 volts when the
*RST command is sent.
The commands are listed in the long form.
The lowercase characters can be omitted.
Sections that are enclosed in braces ([ ])
can also be omitted.
The setting range is listed.
Optional symbols such as m and μ
can also be used.
The unit for the value being set.
The unit can be omitted.
Specify MAX to set the maximum value.
Specify MIN to set the minimum value.
The format of the value that is
returned when a query is sent.
SOUR:DCW:VOLT
Sets the test voltage.
Command
SOURce:DCW:VOLTage[LEVel] {<numeric>
|
MIN
|
MAX}
SOURce:DCW:VOLTage[LEVel]? {MIN
|
MAX}
Parameter
Value:
0
to
6.2k
(The default value is 0.)
Unit:
V
Response
Returns the test voltage in <NR3> format.
The parameters are listed.
In this command, the parameter is numeric. In addition to
specifying the desired value, you can specify the minimum
or maximum value.
* RCL
* RST
Item
Command syntax
Parameters
Units
Queries
Expression format
List of messages
List of errors
Command processing time
See