98
TOS3200
Recalling the Panel Memory and Sequence Program
The PM and STB signals are processed at the times shown below. Check that the READY
signal is low level.
Table 7-2 shows the relationship between the PM0 to PM7 signals and the panel memory
number or program number that is actually recalled.
Fig. 7-3
Strobe signal
Table 7-2
Panel memory and program selection
READY (OUT)
PM0-PM7 (IN)
STB (IN)
10 ms or more
10 ms or more
10 ms or more
5 ms or more
An interval of at least 10 ms is
needed for the next signal input.
MSD
LSD
MANU/AUTO
Recalled data
PM7 PM6 PM5 PM4 PM3 PM2 PM1 PM0
H
H
H
H
H
H
H
H
H
Panel memory 00
L
Program 00
H
H
H
H
H
H
H
L
H
Panel memory 01
L
Program 01
H
H
H
H
H
H
L
H
H
Panel memory 02
L
Program 02
•
•
•
L
H
H
L
L
H
H
H
H
Panel memory 98
L
Program 98
L
H
H
L
L
H
H
L
H
Panel memory 99
L
Program 99
Содержание TOS3200
Страница 15: ...General Description This chapter gives an overview and describes the features...
Страница 20: ...20 TOS3200 Options This page has been intentionally left blank...
Страница 29: ...Panel Control Basics This chapter explains how to select items on the screen and how to enter values...
Страница 38: ...38 TOS3200 Panel Memory This page has been intentionally left blank...
Страница 67: ...Meter Mode Measurement This chapter explains the touch current measurement in Meter Mode Meter Mode Measurement...
Страница 73: ...Program Test This chapter describes how to create and edit sequence pro grams and how to execute the test...
Страница 105: ...Specifications This chapter gives the specifications and external dimensions...
Страница 115: ...Appendix A A List of Default Settings B System Settings C Protection Function D Troubleshooting...
Страница 129: ......