DANGER lamp
• The DANGER lamp illuminates while the test is in progress.
If the DANGER lamp is illuminated, the voltage applied to
AC LINE IN (EUT) is output to the AC LINE OUT terminal.
Measurement category
• Measurement terminals A and B comply with IEC Standard
Measurement Category II.
Measurement Category II (CAT II) is used to measure the
primary circuit of equipment (household electric appliances,
portable tools, etc.) that is connected directly to a low-volt-
age installation such as an outlet.
Ground terminal
• In case of measuring the touch current in the condition of
“Earth line disconnected status” of the EUT, a hazardous
voltage may apply to the ground terminal. Do not touch the
ground terminal while the test is in progress.
N
1500VA
MAX
L
G
DANGER
LEAKAGE CURRENT
AC LINE OUT
LINE
ON
EUT
A
B
CAT II
REMOTE
DANGER lamp
Measurement terminal (CAT II)
Ground
terminal
Rated voltage
Between terminals A and B: 250 Vac
Between the terminal and
the chassis:
250 Vac
Be sure to use the ground
terminal with the binding
post cover.
Содержание TOS3200
Страница 15: ...General Description This chapter gives an overview and describes the features...
Страница 20: ...20 TOS3200 Options This page has been intentionally left blank...
Страница 29: ...Panel Control Basics This chapter explains how to select items on the screen and how to enter values...
Страница 38: ...38 TOS3200 Panel Memory This page has been intentionally left blank...
Страница 67: ...Meter Mode Measurement This chapter explains the touch current measurement in Meter Mode Meter Mode Measurement...
Страница 73: ...Program Test This chapter describes how to create and edit sequence pro grams and how to execute the test...
Страница 105: ...Specifications This chapter gives the specifications and external dimensions...
Страница 115: ...Appendix A A List of Default Settings B System Settings C Protection Function D Troubleshooting...
Страница 129: ......