TOS3200
121
System Settings
Appx
*1
You can hear the buzzer volume by pressing the BUZ CHK (SHIFT+F5) key when BUZ VOL
(PASS) or BUZ VOL (FAIL) is selected.
*2
Even if you set LINE BRK (AUTO) to OFF, the TOS3200 may not supply power to the EUT
depending on the step combination.
Setup Items of SYSTEM2/5
On SYSTEM1/5, press the <NEXT> (F5) key to display SYSTEM2/5.
You can execute a measurement check on SYSTEM2/5. The measurement check verifies
the operation of the current measurement circuit of the TOS3200. For a description of the
measurement check execution, see “Measurement Check.”
Item
Description
Panel operation
4
SELV
Sets the safety extra low voltage (SELV).
If the voltage between measurement terminals A and B exceeds the
SELV, the DANGER lamp will illuminate.
Available in the Meter Mode.
F4 key
Voltage
Selectable range: 10 V to 99 V
Rotary knob
OFF
Disable the SELV check function.
5
BUZ VOL
(PASS)
*1
Sets the buzzer volume for PASS judgments.
SHIFT+F1 key
(Rotary knob)
Level
Selectable range: 0 to 10
6
BUZ VOL
(FAIL)
*1
Sets the buzzer volume for U-FAIL and L-FAIL judgments and CON-
TACT FAIL
SHIFT+F2 key
(Rotary knob)
Level
Selectable range: 0 to 10
7
CONTRAST
Sets the screen brightness.
You can also use the
(SHIFT+
) or
(SHIFT+
) key.
SHIFT+F3 key
(Rotary knob)
Level
Selectable range: 0 to 10
8
LINE BRK
(AUTO)
Sets whether or not to supply power to the EUT during program test
interval time.
SHIFT+F4 key
(rotary knob)
ON
Does not supply power to the EUT during the interval time
OFF
Supplies power to the EUT during the interval time
*2
9
<NEXT>
Displays SYSTEM2/5.
F5 key
See
CHECK
<PREV>
<NEXT>
< MEASURE CHECK >
Push the CHECK button
SYSTEM2 / 5
1
2
3
Item
Description
Panel operation
1
CHECK
Executes the measurement check.
F3 key
2
<PREV>
Displays SYSTEM1/5
F4 key
3
<NEXT>
Displays SYSTEM3/5.
F5 key
Содержание TOS3200
Страница 15: ...General Description This chapter gives an overview and describes the features...
Страница 20: ...20 TOS3200 Options This page has been intentionally left blank...
Страница 29: ...Panel Control Basics This chapter explains how to select items on the screen and how to enter values...
Страница 38: ...38 TOS3200 Panel Memory This page has been intentionally left blank...
Страница 67: ...Meter Mode Measurement This chapter explains the touch current measurement in Meter Mode Meter Mode Measurement...
Страница 73: ...Program Test This chapter describes how to create and edit sequence pro grams and how to execute the test...
Страница 105: ...Specifications This chapter gives the specifications and external dimensions...
Страница 115: ...Appendix A A List of Default Settings B System Settings C Protection Function D Troubleshooting...
Страница 129: ......