4-42 Operation Method
RC04-PCR-LA
4.3.10 Output ON/OFF Phase Setting
Output ON/OFF phase setting is available separately. In the Initial
Setup Status, no output ON/OFF phase is set in the Remote Control-
ler (phase is random). This function is used if output ON/OFF phase
setting is required for a case such as inrush current simulation.
Fig.4-19
Output ON/OFF Phase
Output ON/OFF Phase Setting Procedure
• Output impedance setting is available only in the AC mode.
1. Press the ESC key to call the Home Position.
2. Press the MODE key.
The PCR-LA power supply will enter the key-lock mode automat-
ically.
3. Press the F1 (PHASE) key to select the output ON/OFF
phase setting mode.
4. Press the F3 (OUT_ON) or F5 (OUT_OFF) key to select ON
or OFF phase.
If the F3 (OUT_ON) key is pressed, the following display
appears.
F R E E
1 . 0 0
Ω
O F F
P H A S E
I M P
R E G
F1
F2
F3
F4
F5
P H A S E
F R E E
F R E E
O U T – O N
O U T – O F F
F1
F2
F3
F4
F5
ON phase
OFF phase
Содержание RC04-PCR-LA
Страница 1: ...RC04 PCR LA Part No Z1 002 902 IA002963 Jan 2007 OPERATION MANUAL REMOTE CONTROLLER PCR LA Series...
Страница 10: ...1 4 General RC04 PCR LA...
Страница 78: ...7 4 Specifications RC04 PCR LA 7 2 Dimensions Fig 7 1 Dimension of RC04 PCR LA 40 MAX 55 128 180 Unit mm...
Страница 100: ...A 22 Appendix RC04 PCR LA...