4-28 Operation Method
RC04-PCR-LA
Sequence Execution Mode Operating Procedure
The sequence execution mode allows the items set in the sequence
setting mode to be executed in order. Specify the sequence start
address, end address, and repetition count to execute the sequence.
1. Press the ESC key to call the Home Position.
2. Press the SEQ key.
The PCR-LA power supply will enter the key-lock mode automat-
ically.
3. Press the F4 key to move the cursor to the “END” area on
the upper part of the display, and set a sequence end
address value using the numeric keys or JOG/SHUTTLE.
For setting using the numeric keys, press the ENTER key to con-
fi
rm the value.
Setting is established as long as the start address is less that the
end address. When setting the start and end addresses, always
ensure this relationship is satis
fi
ed.
4. Press the F3 key to move the cursor to the “START” area,
and set a sequence start address value in the same way as
the sequence end address.
0
1
1
R U N
S T R T
E N D
L O O P
F1
F2
F3
F4
F5
(Example of setting the end address to “10”)
0
1 0
1
R U N
S T R T
E N D
L O O P
F1
F2
F3
F4
F5
(Example of setting the start address to “1”)
1
1 0
1
R U N
S T R T
E N D
L O O P
F1
F2
F3
F4
F5