RC04-PCR-LA
Operation Method 4-13
4.3.5 Power Line Abnormality Simulation
This function allows the PCR-LA power supply to simulate a inter-
ruption, fast voltage swell (pops), or fast voltage dip (dips) to conduct
a power line abnormality simulation in the AC mode.
Fig.4-4
Simulated Power Line Abnormality Wave-
forms
• Power line abnormality simulation allows sine waveform out-
put only. Even if a special waveform is selected as described
in “4.3.8 Special Waveform Output”, performing power line
abnormality simulation causes the waveform to be output as a
sine waveform.
Operation of power line abnormality simulation consists of the
parameter setting and execution modes.
The setting mode display consists of three displays that can be
switched using the MENU key. The parameters of the setting display
can be selected using the F1 to F5 keys, and a value is set using the
numeric keys or JOG/SHUTTLE. (For setting using the numeric
keys, press the ENTER key to con
fi
rm the value.)
Fig.4-5
Simulated Waveforms and Parameters
Power interruption
waveform
Fast voltage swell (pop)
Fast voltage dip (dip)
NOTE