IEEE488.2 Common Commands
KIKUSUI Electronics Corp.
PWR-01 Interface Manual
*PSC
Sets whether to clear the event status enable register and the service request en-
able register when the POWER switch is turned on (power-on status).
Command
*PSC <boolean>
*PSC?
Parameter <boolean>
Value:
ON(1)
The *ESE and *SRE settings are cleared when the POWER
switch is turned on.
OFF(0)
The *ESE and *SRE settings are not cleared when the POW-
ER switch is turned on.
(Example) To enable the power-on SRQ function
*PSC 0;*SRE 32;*ESE 128
Response
Returns the power-on status setting in response to *PSC?.
*RST
Resets the panel settings.
Clears alarms (if they cannot be cleared, alarms continue).
Aborts the trigger subsystem operation.
Clears the OPC bit (bit 0) of the status event register.
Setting Items
When *RST performed
OUTPut
OFF
OUTPut:DELay:ON
0.0[s]
OUTPut:DELay:OFF
0.0[s]
OUTPut:EXTernal
OFF
[SOURce:]CURRent
MAXimum
[SOURce:]CURRent:EXTernal:SOURce
NONE
[SOURce:]CURRent:LIMit:AUTO
ON
[SOURce:]CURRent:PROTection
MAXimum
[SOURce:]CURRent:SST:FALL
0.0[s]
[SOURce:]CURRent:SST:RISE
0.0[s]
[SOURce:]CURRent:TRIGgered
MAXimum
[SOURce:]VOLTage
0.0[V]
[SOURce:]VOLTage:EXTernal:SOURce
NONE
[SOURce:]VOLTage:LIMit:AUTO
ON
[SOURce:]VOLTage:LIMit:LOW
MINimum
[SOURce:]VOLTage:PROTection
MAXimum
[SOURce:]VOLTage:SST:FALL
0.0[s]
[SOURce:]VOLTage:SST:RISE
0.0[s]
[SOURce:]VOLTage:TRIGgered
0.0[V]
TRIGger:TRANsient:SOURce
IMMediate
TRIGger:PROGram:SOURce
IMMediate
RES
MINimum
Command
*RST