PLZ-4WL
99
Calibration
6
M
ai
nt
enance
■
Calibration numbers
The numbers that appear in the calibration screen are the calibration numbers.
Alarm
If an alarm occurs during calibration, the PLZ-4WL produces an alarm sound, and the load
turns off. After you remove the cause of the alarm and press ENTER, the alarm sound stops,
and the PLZ-4WL returns to the Calibration screen.
Start calibration again from the calibration number at which the alarm occurred.
Preparation
Before you calibrate the PLZ-4WL, leave it on for 30 minutes or more to warm it up.Warming
up reduces measurement error caused by initial drift. Keep the ambient temperature at 23 ±
5
C.
Equipment used
Connect the equipment as shown below. Select an appropriate shunt resistor according to
type of calibration that you are performing.
1.
CC
(
Low
)
2. CC
(
Mid
)
4. CV 4V
5. CV 30V
PREV
NEXT
3. CC
(
Hi
g
h
)
Calibration
Calibration number
Name
Required Accuracy
Required Rating
DC voltmeter
Within 0.02 %
Measurement voltage
range:
0 V to 35 V
Shunt resistor
0.1 %
For 0.5 A
1
For 1 A
1,2
For 5 A
1
For 10 A
2
For 50 A
1
For 100 A
2
Regulated DC power supply
(constant-voltage power
supply)
–
Voltage: 5 V
Current: 50 A
1
100 A
2
Regulated DC power supply
(constant current power
supply)
–
Voltage: 35 V
Current: 0.5 A
1: PLZ164WL, 2: PLZ334WL
Regulated DC
power supply
Voltmeter
(DVM)
Voltmeter
(DVM)
PLZ-4
W
L
Series
+S
-S
+
I
1
V
1
Shunt resistor
+
–
–
Содержание PLZ164WL
Страница 11: ...Installation and Preparation This chapter describes how to unpack and prepare this product before you use it 1 ...
Страница 25: ...Basic Functions This chapter explains the operations for each operation mode and other basic functions 2 ...
Страница 43: ...Advanced Operations This chapter explains ABC preset memory setup memory protection functions and the menu 3 ...
Страница 57: ...Sequences This chapter explains the sequence function Sequences 4 ...
Страница 79: ...External Control This chapter explains external control 5 ...
Страница 95: ...Maintenance This chapter describes the maintenance procedures including cleaning inspection and calibration 6 ...
Страница 107: ...7 Specifications This chapter contains the specifications and gives the dimensions of the PLZ 4WL ...
Страница 117: ...Appendix A Operating Area B Sequence Program Creation Table C Options ...