PLZ-4W
Applied Operation 6-45
■
Setup procedure for external voltage control
This procedure is common to CC, CR, CP, and CV modes. Follow this procedure for
modes other than CC mode.
• Use external voltage of low noise and high stability.
• Use twisted wires for the signal wires. This can prevent disturbance from noise.
1.
Check that the power is turned off.
2.
Connect the external voltage.
Connect the external voltage across pins 1 and 3 of the J1 connector.
3.
Turn on the power.
4.
Check that the load is turned off.
Check that the LOAD LED is turned off. If the LED is illuminated, press the
LOAD key to cut off the current.
5.
Set the operation mode and current range.
Press the MODE key to select the operation mode.
Press the RANGE key to set the current range. When also controlling the cur-
rent range externally, be sure to select H range on the panel.
6.
Select the menu setup.
Press the MENU (SHIFT+SET/VSET) key.
The the menu appears.
7.
Select Configuration
Press the
or
CURSOR key several times until Configuration is high-
lighted on the menu. When highlighted, press the ENTER key.
8.
Select External.
Press the
or
CURSOR key several times until External is highlighted on
the menu. When highlighted, press the ENTER key.
9.
Select voltage control.
Press the CURSOR key to move the cursor (blinking) to the Control line.
Turn the rotary knob to select V.
10. Exit from the menu.
Press the MENU (SHIFT+SET/VSET) key.
The original screen displayed before entering menu setup appears.
11. Power cycle the PLZ-4W.
The menu settings are confirmed.
• You can prevent the settings from being changed inadvertently during measure-
ment by pressing the LOCK (SHIFT+LOCAL) key to enable the lock.
NOTE
NOTE