PLZ-5W SR Series Setup Guide
7
1
Preparation
This chapter describes how to connect the power cord and how to turn the power on and off.
Connecting the Power Cord
This product conforms to IEC Overvoltage Category II (energy-consuming equipment that is
supplied from a fixed installation).
1
Turn the POWER switch off ( ).
2
Check that the AC power line meets the nominal input rating of the prod-
uct.
The product can receive a nominal line voltage in the range of 100
Vac to 240
Vac at
50
Hz or 60
Hz (frequency range: 47
Hz to 63
Hz).
3
Connect the power cord to the AC INPUT inlet on the rear panel.
WARNING
Risk of electric shock.
• This product conforms to IEC Safety Class I (equipment that has a protective con-
ductor terminal). Be sure to earth ground the product to prevent electric shock.
• The product is grounded through the power cord ground wire. Connect the protec-
tive conductor terminal to earth ground.
• Use the included power cord to connect to the AC line.
If the supplied power cord cannot be used because the rated voltage or the plug shape is
incompatible, have a qualified engineer replace it with an appropriate power cord that is
3 m or less in length. If obtaining a power cord is difficult, contact your Kikusui agent or dis-
tributor.
• Do not use the supplied power cord with other instruments.
• The power cord with a plug can be used to disconnect the product from the AC power line
in an emergency. Connect the plug to an easily accessible power outlet so that the plug can
be removed from the outlet at any time.
• Secure adequate space around the power cord plug. Do not insert the power cord plug into
an outlet that is difficult to access or place objects around the power cord plug that will
make plugging and unplugging difficult.