2
Connecting the EUT
This chapter describes how to connect the EUT to the PLZ-5W SR.
Connecting the Load Cables
Using load cables, connect the EUT to the product through the load input terminals on the
rear panel.
The load input terminals on the front panel are for using a standalone PLZ1205W in a simple
manner. Do not use these terminals as this product uses parallel operation.
Selecting the load cables
Select load cables that are suitable for the current to be used. You can use the optional large-
current load cables. For details on large-current load cables, see the large-current load cable
manual in the included CD-ROM.
WARNING
• Risk of fire. Use load cables whose capacity is adequate for the product’s rated input
current.
• Risk of electric shock. Use load cables with a voltage rating that meets or exceeds
the product’s isolation voltage (±500 Vdc).
CAUTION
• Use load cables with a core diameter that is appropriate for the amount of current being
used and with sturdy, flame-resistant insulation.