24
Operation Manual
PFX2731S
Preparation | Preparing for Connecting a DUT
Constructing output cable
1
Select the appropriate output cable.
Considering the current capacity, we recommend using cables with a cross-sectional area larger
than the nominal value of 5.5 mm
2
.
2
Attach the terminal (pin) to this product end of the output cable and connect it to
the DUT connector.
3
If necessary, attach a compatible crimping terminal to the DUT end of the output
cable.
Connection example of PFX2731S and DUT
Output cable
Voltage sensing cable
(shielded cable)
DUT
Thermocouple
+
–
Temperature measurement box
-
+
-
+
DUT
connector
Up to 5 m in length
Insert here the output cable with the terminals (pins) attached.
Terminal (pin)
Output cable (
−
)
Output cable (
+
)
CAUTION
If the DUT will be subject to high temperature, such as by placing it in a temperature
chamber, pay attention to the allowable temperature of the output cable. If the heat
resistance of the output cable and the like is insufficient, the DUT will be in a dangerous
situation due to poor insulation, increased contact resistance, and so on.
• You can extend the cable length up to 5 m. If the cable is longer than 5 m, constant current
control may become unstable and accurate capacity measurement may be compromised.
• To prevent incorrect wiring, use different color crimping terminals or cables to make the
polarities easily distinguishable.