Combination for Performing Charge/Discharge Tests
Selecting the appropriate DC power supplies and electronic
loads
The range is different between charging and discharging. Select DC power supplies and electronic
loads that meet each condition.
The graph below shows the range of chargeable voltage and current using the following equation.
Use devices whose voltage (constant voltage, cutoff voltage) and current (constant current) do not
exceed the range.
Maximum charge power = Maximum rated power of DC power supply – Loss in cables
1
1
What is loss in cables?
Voltage drops occur as a result of charge currents flowing through DUT cables, connection
cables, PFX2532 current path circuits, and so on. Loss in cables is the power loss during
charging due to this voltage drop. The maximum power that can be used for charging is the
resultant value after subtracting the loss in cables.
If the allowable power of DC power supplies is exceeded during charging, PS / B alarm or the like
occurs, and the test is aborted.
Voltage (V) axis
Constant voltage value (CV Voltage) or cutoff voltage value (Cutoff voltage)
Current (A) axis
Constant current
value (CC Current)
0
10
20
20
40
60
80
100
120
140
160
180
200
30
40
50
60
70
PWR1600Lx2
PAT60-133T
PAT40-200T
PWR1201L
Voltage (V)
Charge operating range of the each DC power supply
[PAT-T/ PWR-L/ PWR-01L]
DUT cable: At nominal cross-sectional area 80 mm
2
3 m
PFX2532
operating range
Current (A)
Содержание PFX2532
Страница 12: ...12 PFX2532 This page is intentionally blank ...
Страница 25: ...Connecting and Configuring Devices This chapter describes how to connect and configure the devices ...
Страница 62: ...62 PFX2532 This page is intentionally blank ...
Страница 71: ...Specifications This chapter provides the PFX2532 s specifications and outline drawing ...
Страница 96: ...96 PFX2532 This page is intentionally blank ...
Страница 141: ......