PCZ1000A
3-31
3
Ope
ration
Calibration Procedure
If calibration is interrupted before being finished, the next calibration will be
restarted from the same step where it left off.
1
Press the LOAD key to set the instrument to the load-off state.
2
Press SHIFT + I SET (CONFIG).
If "Norm/M/S" is not displayed, press the ENTER key to select ">3: Norm/M/S".
3
Use the rotary knob to select "MASTER".
4
Press the ENTER key to select ">2:Para/Track".
5
Use the rotary knob to select "PARALLEL".
6
Press the ENTER key to select ">3: Parallel Ope".
7
Use the rotary knob to set the "number of units".(This example sets the
number of units to two.)
8
Press the ENTER key to display ">4: Calibration".
This changes the PCZ1000A to calibration mode. Proceed to the calibration procedure.
■
Calibrating the offset and gain for the total current (RMS)
1
Press the I SET key to display "IRS OFFS XXXXX".
XXXXX is the RMS offset value.
2
Press the LOAD key to change the instrument to the load-on state.
3
Use the rotary knob to set the offset value (variable range:
-7 000 to 1 000) on the front panel display to the value on the externally
connected AC ammeter (RMS meter).
> 1: Norm/M/S
MASTER
> 2: Para/Track
PARALLEL
> 3: Parallel Ope.
2
Содержание PCZ1000A
Страница 4: ...ii PCZ1000A This page is intentionally blank ...
Страница 14: ...xii PCZ1000A Front panel 18 17 Control panel Control panel 1 Display Display 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Страница 17: ...General Description This chapter gives an overview and describes the features of the PCZ1000A ...
Страница 35: ...Operation This chapter describes each mode and function of the operation ...
Страница 82: ...5 4 PCZ1000A This page is intentionally blank ...
Страница 83: ...Specifications This chapter gives the specifications and dimensions of the PCZ1000A ...
Страница 88: ...6 6 PCZ1000A This page is intentionally blank ...
Страница 89: ...The appendix contains the basic operating modes operating range and troubleshooting Appendix ...
Страница 99: ......