PCZ1000A
3-19
3
Ope
ration
Changing Preset Memory
The following example explains how to change the range for the limit value saved
in preset memory A for CR mode and how to change the limit value saved in preset
memory C.
1
Press the R SET key.
The resistance limit value and range are displayed in CR mode.
2
Press the SHIFT + MEM (DMEM) key.
The settings are stored in preset memory A. Pressing the A key displays "Rval" in the
lower part of the display (the letter enclosed in "< >" indicates the preset memory
location.)
3
Press the A key.
4
Press the SHIFT + RSET (RANGE) key.
This changes the range from "H" to "L". The range changes to the new setting
immediately.
5
Press the SHIFT + MEM (DMEM) key or ESC key to end the operation.
6
Press the SHIFT + R SET (RANGE) key.
7
Press the C key.
8
Use the rotary knob to change the limit value.
The limit value changes to the new setting immediately.
9
Press the SHIFT + MEM (DMEM) key or ESC key to end the operation.
The display returns to its state prior to the change.
7.13Ar10.1Ap 84.9V
--RSET 11.111
·
H--
7.13Ar10.1Ap 84.9V
--Rval<A>10.000
·
H--
7.13Ar10.1Ap 84.9V
--Rval<A>10.000
·
L--
7.13Ar10.1Ap 84.9V
--Rval<C>9.0909
·
H--
Содержание PCZ1000A
Страница 4: ...ii PCZ1000A This page is intentionally blank ...
Страница 14: ...xii PCZ1000A Front panel 18 17 Control panel Control panel 1 Display Display 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Страница 17: ...General Description This chapter gives an overview and describes the features of the PCZ1000A ...
Страница 35: ...Operation This chapter describes each mode and function of the operation ...
Страница 82: ...5 4 PCZ1000A This page is intentionally blank ...
Страница 83: ...Specifications This chapter gives the specifications and dimensions of the PCZ1000A ...
Страница 88: ...6 6 PCZ1000A This page is intentionally blank ...
Страница 89: ...The appendix contains the basic operating modes operating range and troubleshooting Appendix ...
Страница 99: ......