KIKUSUI Electronics Corp.
90
INSTrument Command
This command selects the phase that SOURce and MEASure/FETCh commands
apply to for single-phase three-wire output or three-phase output. The INSTrument
command designation does not apply to phase-designated commands.
In the factory default condition, at power-on, and after a *RST or *RCL is sent, all
phases are selected (INST:COUP ALL).
INST/ INST:NSEL
Selects the phase that SOURce and MEASure/FETCh commands apply to.
This does not apply to phase-designated commands.
This is invalid when INST:COUP is set to ALL.
Command
INSTrument[:SELect] <character>
INSTrument[:SELect]?
INSTrument:NSELect <NRf>
INSTrument:NSELect?
Parameter
INST command
Value:
OUTPut1
U phase (default)
OUTPut2
V phase
OUTPut3
W phase
Parameter
INST:NSEL command
Value:
1
U phase (default)
2
V phase
3
W phase
Settings are reset to default values when an *RST command is sent.
Example
INST OUTP1
Response
Returns the designated phase in character format in response to INST?.
Returns the designated phase in NR1 format in response to INST:NSEL?.