TRIGger Command
PCR-WE Interface Manual
299
TRIG:TRAN:SOUR
Sets the condition (trigger source) for actually changing the settings after the TRAN-
sient trigger subsystem receives an INIT:TRAN.
Command
TRIGger:TRANsient:SOURce <character>
TRIGger:TRANsient:SOURce?
Parameter
Value:
IMMediate
Change the settings immediately (default)
BUS
Change the settings when a software trigger (*TRG,
TRIG:TRAN, IEEE488.1 get (Group Execute Trigger) is re-
ceived
Settings are reset to default values when an *RST command is sent.
Example
TRIG:TRAN:SOUR BUS
Response: Characters