background image

1

2

3

4

7

5

8

6

9

10

12

15

21

22

23

24

25

16

17

11

13

14

18

19

19

20

PCR-LE  PCR-LE2

Quick Reference

Z1-006-102, IB027724

Sep. 2017

The newest version of the 

operation manual can be 

downloaded from Kikusui website.

http://www.kikusui.co.jp/en

Name

Function

1

OUTPUT key

Turn output on and off

2

Display

Display the settings/ measured value

3

Numeric keypad

Enter numeric values directly

4

ON PHASE key

Set the output on phase

5

RANGE key

Switch output voltage range

6

SYNC key

Set the synchronization function

7

OFF PHASE key

Set the output off phase

8

KEYLOCK key

Lock and unlock the keys

9

RESET key

Reset the product

10

SLEEP key

Set the sleep mode

11

LOCAL key

Switch between remote and local mode

12

+/- key

Switch the polarity of the voltage

13

DIGIT key

The cursor moves to the left

14

CLR key

Clear the numeric keypad input

ALM CLR key

Clear alarms

15

F key

Set the frequency

PHASE key

Switch different displays during multiphase 
operation (optional)

Name

Function

16

I key

Configure the current settings

I/P/

λ

-MEAS key

Switch the current/power measurement mode

17

V key

Configure the voltage settings

V-MEAS key

Switch the voltage measurement mode

18

ESC key

Cancel operations

19

Detachment button Detach the control panel

20

Function key

Correspond to the menus on display

21

MEMORY key

Save/Load settings from memory

OHTERS key

Configure advanced operation settings

22

SHIFT key

Enable the functions that are indicated in the 
bottom row to the left of each key

23

ENT key

Apply settings

24

SIM key

Configure power line abnormality simulations

SEQ key

Configure sequence operations

25

OPR MODE key

Configure the operation environment settings

CONFIG key

Specify the configuration settings

Main specifications

1P: Single-phase input/output, 1P3W: Single-phase three-wire output, 3P: Three-phase output, 3P3W: Three-phase three-wire input, 3P4W: Three-phase 
four-wire input 
TYP: Typical values that are representative of situations where the PCR-LE/PCR-LE2 Series operates in an environment with an ambient temperature of 23 °C.

Item

PCR
500LE

PCR
1000LE

PCR
2000LE

PCR
3000LE

PCR
4000LE

PCR
6000LE

PCR
9000LE

PCR
6000LE2

PCR
9000LE2

PCR

12000LE2

PCR

18000LE2

PCR

27000LE2

Input ratings (AC rms)
Voltage

1P

100 V input type: 85 Vac to 132 Vac, 
200 V input type: 170 Vac to 250 Vac

170 Vac to 

250 Vac

--

170 Vac to 

250 Vac

--

3P3W

--

170 Vac to 250 Vac

3P4W

--

187 Vac to 250 Vac  (Phase voltage)

Frequency

47 Hz to 63 Hz

Apparent power

Approx. 
0.93 kVA

Approx. 
1.8 kVA

Approx. 
3.6 kVA

Approx. 
5.5 kVA

Approx. 
7.3 kVA

Approx. 
10.6 kVA

Approx. 
15.7 kVA

Approx. 
10.6 kVA

Approx. 
15.7 kVA

Approx.  
23 kVA

Approx.  
33 kVA

Approx.  
48 kVA

AC mode output ratings (AC rms)
Voltage

Output L range: 1 V to 150 V, Output H range: 2 V to 300 V

Resolution

0.1 V

Maximum 
current(Output 
L range, H 
range) 

*1

1P

5 A, 
2.5 A

10 A, 5 
A

20 A, 
10 A

30 A, 
15 A

40 A, 
20 A

60 A, 
30 A

90 A, 
45 A

60 A, 
30 A

90 A,  
45 A

120 A,  
60 A

180 A,  
90 A

270 A,  
135 A

1P3W, 3P ---

20 A, 
10 A

30 A, 
15 A

40 A, 
20 A

60 A, 
30 A

90 A, 
45 A

Power capacity 1P

500 VA

1 kVA

2 kVA

3 kVA

4 kVA

6 kVA

9 kVA

6 kVA

9 kVA

12 kVA

18 kVA

27 kVA

1P3W

---

4 kVA

6 kVA

8 kVA

12 kVA

18 kVA

3P

6 kVA

9 kVA

12 kVA

18 kVA

27 kVA

Maximum peak current

Max. current (rms) × 4 (TYP)

*2

Maximum reverse current 

*3

30 % of the max. current (rms)

Frequency

*1

1 Hz to 999.9 Hz

*4

Resolution

0.01 Hz (1.00 Hz to 100.0 Hz), 0.1 Hz (100.0 Hz to 999.9 Hz)

DC mode output ratings
Voltage (output L range, H 
range)

1.4 V to 212 V, 2.8 V to 424 V

Resolution

0.1 V

Maximum 
current(Output 
L range, H 
range)

*5

1P

3.5 A, 
1.75 A

7 A, 
3.5 A

14 A, 
7 A

21 A, 
10.5 A

28 A, 
14 A

42 A, 
21 A

63 A, 
31.5 A

42 A, 
21 A

63 A,  
31.5 A

84 A, 
42 A

126 A,  
63 A

189 A, 
94.5 A

1P3W

---

14 A, 
7 A

21 A,  
10.5 A

28 A, 
14 A

42 A, 
21 A

63 A,  
31.5 A

Maximum instantaneous 
current

Max. current (rms) × 3.6 (Limited by the rated output current’s rms value)

Power capacity 1P

350 W

700 W

1.4 kW

2.1 kW

2.8 kW

4.2 kW

6.3 kW

4.2 kW

6.3 kW

8.4 kW

12.6 kW 18.9 kW

1P3W

---

2.8 kW

4.2 kW

5.6 kW

8.4 kW

12.6 kW

Output voltage waveform distortion ratio, Output voltage response speed
Output voltage 
waveform 
distortion 
ratio

*6

FAST

0.2 % or less

--

MEDIUM

0.3 % or less

0.5 % or less

Output voltage 
response 
speed  

*7

FAST

20 μs (TYP)

--

MEDIUM

30 μs (TYP)

50 μs (TYP)

General
Input terminal

Inlet

M4

M5

M8

M8

M8

*8

M5

M8

*8

M5

M8

M8

M8

Output terminal

M4

M4

M4

M5

M5

M8

M8

M8

*9

M8

*9

M8

M8

M8

*1

. When the output voltage is between 1 V and 100 V or 2 V and 200 V and the load power factor is between 0.8 and 1. When the output voltage is 

between 100 V and 150 V or 200 V and 300 V, the output current is reduced by the output voltage. When the load power factor is between 0 and 0.8, the 
output current is reduced by the load power factor. When the output frequency is between 1 Hz and 40 Hz, the output current is reduced by the output 
frequency.  

*2

. For capacitor-input rectifier loads, (at near the peak of the voltage waveform, excluding three-phase three-wire output). The peak current 

that can be output decreases in accordance with the reduction in the absolute value of the instantaneous output voltage. 

*3

. When the output voltage is 

100 V or 200 V and the output frequency is between 40 Hz and 999.9 Hz (reverse current is -180 deg out of phase with the output voltage). 

*4

. 1 Hz to 500 

Hz when a 3P05-PCR-LE (500HZ LMT) is installed in the PCR-LE and on the PCR-LE2 500HZ LMT model (during three-phase output). 

*5

. When the output 

voltage is between 100 V and 212 V or 200 V and 424 V, the output current is reduced by the output voltage. 

*6

. When the output voltage is between 80 

V and 150 V or 160 V and 300 V and the load power factor is 1. 

*7

. When the output voltage is 100 V or 200 V, the load power factor is 1, and the output 

current changes from 0 A to the rated value and from the rated value to 0 A. 

*8

. Three-phase three-wire or Three-phase four-wire input: M5 

*9

. Single-phase 

three-wire or Three-phase input: M5

The accompanying CD-ROM describes the following 

information.

Two-phase output setting PCR-
LE optional
Voltage compensation function
Harmonic current analysis 
function
Generating special waveforms 
(waveform bank)
Setting the output impedance
Setting soft starts (the voltage 
rise time)
Fixing the internal Vcc
Selecting the response
Controlling the Output Using 

External Analog Signals 
(Optional)
Remote control
Sequence tutorial
Options
Troubleshooting
Maintenance
Factory default settings
Alarms/ Trouble
Specifications
Table for power line abnormality 
simulation settings
Table for sequence settings

MANUAL

Alarms/Trouble

When a protection function is activated, an alarm (ALM-xx) or a 
trouble indication (TRBL-xx) is generated, and the output is turned 
off.
Alarm: Press ALM CLR(SHIFT+CLR) to clear the alarm, and then fix 
the problem that caused the alarm.
Trouble: To clear the trouble indication, turn the POWER switch off, 
wait at least 5 seconds, and turn it back on. If the same trouble 
persists, contact your Kikusui agent or distributor to request repairs.

PCR-LE 

manual

PCR-LE2 

manual

Отзывы: