PBZ
87
CONFIG Setting Details
3
Memory
, Response, an
d Config
uratio
n
CONFIG[2]: Signal Source, External Signal Circuit Gain, and
the Power-On State
• The PBZ does not change automatically from CV mode to CC mode.
• You cannot change the following settings when the output is on.
• CONFIG[1]>MODE>POLARITY
• CONFIG[1]>MODE>CONTROL
• CONFIG[1]>MODE>SENSING
SIGNAL SOURCE
SELECT
INT
EXT
ADD
EXT SELECT
BNC
J1
BNC+J1
EXT GAIN (PBZ20-20)
EXT GAIN (PBZ40-10)
CV: 0.0 to ±20.0
CV: 0.0 to ±40.0
CC: 0.0 to ±20.0
CC: 0.0 to ±10.0
POWER ON
STATE
DEFAULT
LAST
OUTPUT
OFF
ON
SEQ RUN
OFF
PROG1 to PROG16
CV SCRIPT
CC SCRIPT
Item
Description
Value
Actual value
Factory default
setting
SIGNAL SOURCE
>SELECT
Signal source.
INT
Internal signal
source.
INT
EXT
External signal
source.
ADD
INT and EXT are
added.