A-26
PAT-T 4kW
A.6.5
Error Checking
Error/event queue
The SCPI specifications define a standard error reporting scheme, Error/Event
Queue. This is a FIFO (First In First Out) queue, which records errors and events.
The maximum number of errors/events that the PAT can record is 255. Each error/
event can be read with the SYSTem:ERRor query.
:SYSTem:ERRor?
The response to this query contains a numeric part (error/event number) and a
textual description, such as:
-222,"Data out of range"
The error/event queue is empty when the *CLS common command is sent, when the
last item in the queue is read, or when the PAT is turned on. When the error/event
queue is empty, the query returns the following:
0,"No error"
Displaying communication errors
The PAT has a debug trace function. The oldest item among the errors and events (if
they are present) can be displayed on the PAT. This function is convenient when you
debug your remote applications.
While an error/event item is displayed on the panel, the normal voltmeter and
ammeter are disabled.
If the error/event queue is empty, the debug trace function does not display any
errors. Sending the *CLS command clears the communication error display.
If in local mode, the debug trace function is temporarily disabled.
The communication error display can be enabled or disabled with the
SYSTem:CONFigure:TRACe command.
SYSTem:CONFigure:TRACe <ON|OFF>
Содержание PAT160-25T
Страница 17: ...General Description This chapter describes an overview and describes the features...
Страница 22: ...1 6 PAT T 4kW...
Страница 34: ...2 12 PAT T 4kW...
Страница 73: ...External Control This chapter describes external analog control and remote monitoring using the J1 connector...
Страница 106: ...6 12 PAT T 4kW...
Страница 158: ...7 52 PAT T 4KW...
Страница 159: ...Maintenance This chapter describes maintenance such as cleaning inspecting and calibrating...
Страница 171: ...Specifications This chapter describes the specifications and gives the dimensions of the PAT...