SYST:CONF:BTR:PROT
Sets whether to trip the breaker (turn the POWER switch off) when the overvoltage
protection (OVP) or overcurrent protection (OCP) is activated.
Command
SYSTem:CONFigure:BTRip:PROTection {ON|OFF|1|0}
SYSTem:CONFigure:BTRip:PROTection?
Parameter
Value ON(1) Trip (turn the POWER switch off)
OFF(0) Not trip (turn the output off) (default)
Response
Returns whether to trip the breaker when the OVP or OCP is activated in the NR1 form in
response to SYST:CONF:BTR:PROT?.
Overview
Setup
Overview of Messages
Command
(function search)
Protection and Clear the Alarm
Setting Changes using triggers
Command
(ABC search)
Command
(Sub-system search)
Appendix
Tutorial
PAT-T series Communication Interface Manual
KIKUSUI ELECTRONICS CORP.
91/158