FETC:VOLT
If FETC? is sent without making a single measurement after resetting the PAT-T (using
*RST) or after switching from local mode (panel operation) to remote mode, an SCPI error
(-230,"Data corrupt or stale") occurs. This is because there is no valid measured data that is
held. Be sure to use this command after the
or MEAS command.
Command
FETCh[:SCALar]:VOLTage[:DC]?
Response
Returns the measured value of the voltage output in the NR3 form in response to
FETC:VOLT?.
If the measurement is not complete, the response data is generated after the measurement
for the FETC:VOLT query is complete.