PAS SERIES
Basic Operation 3-11
3.2.5 Using the Unit as a Constant Voltage Power
Supply
This section describes the procedure for using the unit as a constant
voltage power supply.
1.
Check that the POWER switch is turned OFF.
2.
Connect the load to the output terminal.
For details on connecting the load, see "3.3 Connecting the Load".
3.
Turn on the POWER switch.
4.
Check that the OUTPUT OFF is illuminated.
5.
Press the SET switch to enter the voltage/current setup
mode.
6.
Press the CURRENT switch to select Coarse or Fine for
setting the current.
7.
Turn the dial to set the value of the current that can flow to
the load.
The value entered here becomes the current limit.
8.
Press the VOLTAGE switch to select Coarse or Fine for set-
ting the voltage.
9.
Turn the dial to set the desired voltage.
10.
Press the OUTPUT switch to illuminate OUTPUT ON.
11.
Voltage is output to the output terminal.
OUTPUT ON and CV light up, and the voltage display shows the
actual output voltage.
You can set the voltage while checking the actual output voltage even
when OUTPUT ON is illuminated.
The current limit must be changed by pressing the SET switch.
• If the output current exceeds the current limit that was speci-
fied in step 7 due to load fluctuations when the unit is oper-
ating in constant voltage mode, the unit switches to constant
current mode. When the unit switches to constant current
mode, CC lights up.
NOTE
Содержание PAS 10-105
Страница 10: ...VIII PAS SERIES This page has been intentionally left blank ...
Страница 54: ...3 22 Basic Operation PAS SERIES This page has been intentionally left blank ...
Страница 78: ...4 24 Remote Control PAS SERIES This page has been intentionally left blank ...
Страница 131: ...PAS SERIES Specifications 8 13 This page has been intentionally left blank ...
Страница 137: ...PAS SERIES Specifications 8 19 This page has been intentionally left blank ...
Страница 145: ......