74
KHA3000
Power-On of the Test System
This section explains how to turn on the power to the test system including this product, the
AC Power Supply and Line Impedance Network. There are the following two methods for
turning on/off the power switch:
●
Separate ON/OFF
Turn on and off the POWER switches of individual devices separately.
●
Collective ON/OFF
Turn on and off the power collectively using an external power switch. Lock the POWER
switches of individual devices in the ON position. The customer is requested to prepare
the external power switch.
Separate ON/OFF
This method is normally used to control the AC Power Supply from this product. This
procedure is explained below.
To prevent damage to the internal relay contact of the Line Impedance Network, follow the
procedure exactly. For details on preventing damage to the internal relay contact, refer to the
Operation Manual of the Line Impedance Network.
Turning on the POWER switch
1
Turn on the POWER switch of the Line Impedance Network.
2
Turn on the POWER switch of the AC Power Supply.
The version is displayed for several seconds on the control panel of the AC Power
Supply. The output remains at OFF. For details on the version display, refer to the
Operation Manual of the AC Power Supply.
Subsequent operations are controlled from this product, KHA3000.
3
Turn on the POWER switch of this product.
Press the POWER switch ( ). With the startup sound, all LEDs illuminate and the
firmware version is displayed.
Содержание KHA3000
Страница 20: ...20 KHA3000...
Страница 35: ...Overview This chapter provides an overview of this product and explains the options for the product 1...
Страница 42: ...42 KHA3000 Options...
Страница 49: ...Wiring Measurement Circuits This chapter explains the wiring of measurement circuits and the control circuit 3...
Страница 70: ...70 KHA3000 Connecting Communication Cables...
Страница 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Страница 91: ...Setting the Harmonic Test This chapter explains how to set the test conditions for the harmonic current test 5...
Страница 120: ...120 KHA3000 Controlling the AC Power Supply...
Страница 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Страница 247: ...Other Measurements This chapter explains basic measurement FFT analyzer and in rush current measurement 11...
Страница 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Страница 352: ...352 KHA3000 Cleaning and Checking...
Страница 353: ...Specifications This chapter explains the specifications of this product 15...
Страница 390: ...390 KHA3000...
Страница 391: ......