54
KHA3000
Wiring Measurement Circuits
This section explains how to wire single-phase two-wire, single-phase three-wire, three-phase
three-wire and three-phase four-wire circuits. An impedance network is used for a flicker test.
Because the impedance network is not used for a harmonic current test except for some special
cases, connect the AC power supply directly to the SOURCE terminals of this product.
Input phases (channels) are L1, L2 and L3. L1 is the reference phase for measurement. Be
sure to wire L1.
Circuit Diagrams
This section shows individual types of measurement circuits. For actual wiring, refer to the
connection table and wiring diagram in the section explaining "Wiring Measurement Circuits".
Single-phase two-wire circuit
WARNING
•
This work has a risk of electric shock. For the AC power supply used for the
measurement circuit, be sure to disconnect the power cord from the outlet, turn off
the switch on the switchboard to which the power cord is connected, and turn off the
POWER switch.
CAUTION
Terminal G of this product is not a protective conductor terminal. It is not grounded even if
the wire is connected to terminal G. For terminals G and N, the respective SOURCE and
LOAD terminals are short-circuited internally.
See
A
V
L
N
G
L
N
G
N
G
L1
AC power
supply
Impedance
network
KHA3000
EUT
SOURCE
LOAD
Содержание KHA3000
Страница 20: ...20 KHA3000...
Страница 35: ...Overview This chapter provides an overview of this product and explains the options for the product 1...
Страница 42: ...42 KHA3000 Options...
Страница 49: ...Wiring Measurement Circuits This chapter explains the wiring of measurement circuits and the control circuit 3...
Страница 70: ...70 KHA3000 Connecting Communication Cables...
Страница 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Страница 91: ...Setting the Harmonic Test This chapter explains how to set the test conditions for the harmonic current test 5...
Страница 120: ...120 KHA3000 Controlling the AC Power Supply...
Страница 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Страница 247: ...Other Measurements This chapter explains basic measurement FFT analyzer and in rush current measurement 11...
Страница 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Страница 352: ...352 KHA3000 Cleaning and Checking...
Страница 353: ...Specifications This chapter explains the specifications of this product 15...
Страница 390: ...390 KHA3000...
Страница 391: ......