3
KHA3000
53
Common Information on Measurement Circuits
W
iring M
easurem
ent Circuits
Mechanism of Voltage Measurement
The figure below shows voltage terminal switching circuits. In the single-phase two-wire,
single-phase three-wire and three-phase four-wire circuits, the switch provided on one end of
each voltmeter connects the voltmeter of each phase between L1 and N, L2 and N, or L3 and
N.
In the three-phase three-wire circuit, the voltmeter of each phase is connected between L1
and L2, L2 and L3, or L3 or L1.
WARNING
•
If the input voltage exceeds 400 Vrms, the primary circuit of equipment connected
directly to an outlet (CAT II) cannot be measured. Only the circuit that is not directly
connected to a commercial power supply (CAT I) is measured.
•
With the input voltage equal to or less than 400 Vrms, the circuit that is not directly
connected to a commercial power supply (CAT I) and the primary circuit of
equipment connected directly to an outlet (CAT II) can be measured.
Item
Performance
Remarks
Maximum input voltage 600 Vrms/900 Vpeak
Voltage input terminals and current input terminals
between L1 and N, L2 and N, L3 and N, L1 and
L2, L2 and L3, and L3 and L1
Measurement category CAT I
*1
between 500
Vrms (exclusive) and 600 Vrms
Measurement category CAT II
*2
below 400 Vrms
*1.
Used to measure circuits that are not directly connected to a commercial power supply
*2.
Used to measure the primary circuit of equipment (home appliances, portable tools, etc.) that
is directly connected to a low voltage facility such as outlets
Maximum input current
40 Arms/100 Apeak
Current input terminal
Instrument loss (input
resistance) of voltage
measurement
About 4.7 M
Voltage input terminals between L1 and N, L2 and
N, L3 and N, L1 and L2, L2 and L3, and L3 and L1
Instrument loss (input
resistance) of voltage
measurement
About 3.5 m
Current input terminal
A
V
A
V
A
V
G
N
V
L1
L2
L3
V
V
V
SOURCE
LOAD
KHA3000
VOLTAGE
SENSING/INPUT
A
V
L1
L2
L3
N
න⋧㧞✢
න⋧㧟✢
ਃ⋧㧠✢
3P3W
1P2W
1P3W
3P4W
Содержание KHA3000
Страница 20: ...20 KHA3000...
Страница 35: ...Overview This chapter provides an overview of this product and explains the options for the product 1...
Страница 42: ...42 KHA3000 Options...
Страница 49: ...Wiring Measurement Circuits This chapter explains the wiring of measurement circuits and the control circuit 3...
Страница 70: ...70 KHA3000 Connecting Communication Cables...
Страница 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Страница 91: ...Setting the Harmonic Test This chapter explains how to set the test conditions for the harmonic current test 5...
Страница 120: ...120 KHA3000 Controlling the AC Power Supply...
Страница 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Страница 247: ...Other Measurements This chapter explains basic measurement FFT analyzer and in rush current measurement 11...
Страница 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Страница 352: ...352 KHA3000 Cleaning and Checking...
Страница 353: ...Specifications This chapter explains the specifications of this product 15...
Страница 390: ...390 KHA3000...
Страница 391: ......