Appx
KHA3000
385
F
Troubleshooting
The following explains check items and measures for various problems. Typical failure symptoms
are described. If a problem occurs, use the following tables to find an appropriate corrective action.
Some problems can be easily solved.
If the problem were not applied to the corresponding item in the table, we recommend you to
reset the product to the default settings. If the problem persists, contact a KIKUSUI agent or
distributor.
Power-on failure
Panel operation failure
See
Symptom
Check item/Corrective action
Page
The POWER switch does not
work.
•
Is the power cord correctly connected?
•
Is the power fuse not blown?
See
Symptom
Check item/Corrective action
Page
The screen display is difficult
to read.
•
Is the screen brightness setting appropriate?
Panel key operation does not
work.
•
Is the key icon on the upper part of the screen LOCK?
Release the key lock.
•
Is the REMOTE LED lit
The RS232C, GPIB or USB interface is enabled.
To enable panel operation, press the LOCAL key to enter the
local mode.
Menu operation does not
work.
•
Did you enter another test mode during menu operation?
•
Is the
“
It cannot be operated from sub-menu
”
dialog box
displayed?
Press the
“
Return
”
function key to exit from the sub-menu.
The test does not start even
when the START switch is
pressed.
•
Did you terminate the previous test in the test ending menu?
When the measurement time is finished, the system enters the
analysis state. In this state, you cannot start a test. Terminate
the test in the test ending menu.
See
Содержание KHA3000
Страница 20: ...20 KHA3000...
Страница 35: ...Overview This chapter provides an overview of this product and explains the options for the product 1...
Страница 42: ...42 KHA3000 Options...
Страница 49: ...Wiring Measurement Circuits This chapter explains the wiring of measurement circuits and the control circuit 3...
Страница 70: ...70 KHA3000 Connecting Communication Cables...
Страница 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Страница 91: ...Setting the Harmonic Test This chapter explains how to set the test conditions for the harmonic current test 5...
Страница 120: ...120 KHA3000 Controlling the AC Power Supply...
Страница 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Страница 247: ...Other Measurements This chapter explains basic measurement FFT analyzer and in rush current measurement 11...
Страница 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Страница 352: ...352 KHA3000 Cleaning and Checking...
Страница 353: ...Specifications This chapter explains the specifications of this product 15...
Страница 390: ...390 KHA3000...
Страница 391: ......