15
KHA3000
361
Specifications
Specifications
External Equipment Control Function
General Specifications
Item
Specifications
AC power supply
control
Setup item
Voltage, frequency, range, ON phase, OUTPUT on/off
RS232C
communication
speed
9600 bps
Item
Specifications
AC input
Nominal input rating
100 V to 240 V, 50 Hz to 60 Hz
Voltage range
90 V to 250 V
Maximum power
consumption
250 VA
Withstanding
voltage
AC input
chassis
No abnormality should occur at 1500 Vac for one minute.
Test terminal
chassis
No abnormality should occur at 3550 Vac for one minute.
Test terminal
Test terminal
Insulation
resistance
AC input
chassis
500 Vdc, 100 M
or higher
Test terminal
chassis
Earth continuity Power code
connector, grounding
pin
chassis
25 Aac/0.1
or less
Environment
Installation location
Indoor, up to 2000 m, overvoltage category II
Temperature
and humidity
ranges
Operating
temperature
0
°C
to +40
°C
(+32
°F to +104 °F
)
Operating humidity
20 %rh to 80 %rh (no condensation)
Storage temperature
-20
°C
to +70
°C (-4 °F to +158 °F)
Storage humidity
90 %rh or less (no condensation)
Safety
Conforming to the following standard requirements:
IEC 61010-1:2001
(Class I
*1
, Pollution degree 2
*2
)
*1.
This is a Class I equipment. Be sure to ground this product’s protective conductor terminal. The
safety of this product is only guaranteed when the product is properly grounded.
*2.
Pollution is addition of foreign matter (solid, liquid or gaseous) that may produce a reduction of
dielectric strength or surface resistivity. Pollution Degree 2 assumes that only non-conductive
pollution will occur except for an occasional temporary conductivity caused by condensation.
Outside dimensions
Refer to
Weight
Approx. 10 kg (22.05 lb)
Accessories
Power cord
1
Voltage sensing
terminal plug
3 (product shipped with this plug installed)
Voltage sensing
terminal short-circuit
wire kit
3 (two wires: product shipped with these wires attached to the
voltage sensing terminal)
Voltage sensing terminal
plug screwdriver
1
Spare fuse
1 (250 V, 2.5 A(T), in the power supply fuse holder)
Operation manual
1 (this manual)
Содержание KHA3000
Страница 20: ...20 KHA3000...
Страница 35: ...Overview This chapter provides an overview of this product and explains the options for the product 1...
Страница 42: ...42 KHA3000 Options...
Страница 49: ...Wiring Measurement Circuits This chapter explains the wiring of measurement circuits and the control circuit 3...
Страница 70: ...70 KHA3000 Connecting Communication Cables...
Страница 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Страница 91: ...Setting the Harmonic Test This chapter explains how to set the test conditions for the harmonic current test 5...
Страница 120: ...120 KHA3000 Controlling the AC Power Supply...
Страница 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Страница 247: ...Other Measurements This chapter explains basic measurement FFT analyzer and in rush current measurement 11...
Страница 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Страница 352: ...352 KHA3000 Cleaning and Checking...
Страница 353: ...Specifications This chapter explains the specifications of this product 15...
Страница 390: ...390 KHA3000...
Страница 391: ......