13
KHA3000
343
SCPI Register Model
Rem
ote Contr
ol
QUEStionable Status Register (STATus:QUEStionable)
The QUEStionable status register is a 16-bit register containing information on questionable
events and status generated during operation of this product.
Some register bits indicate that there is a problem in the measurement data of this product.
STAT:QUES
Queries the event in the QUEStionable status register.
The query clears the register data.
Command
STATus:QUEStionable[:EVENt]?
Response
Returns the event in the QUEStionable status register in the <NR1> format.
STAT:QUES:COND
Queries the status of the QUEStionable status register.
The query does not clear the register data.
Command
STATus:QUEStionable:CONDition?
Response
Returns the status of the QUEStionable status register in the <NR1> format.
Bit
Bit
weight
Bit name
Description
0
1
Over-range on Vm
Voltage measurement exceeded the range.
1
2
Over-range on Cm
Current measurement exceeded the range.
2
4
NPOW
The total real power is negative.
3
8
CLAMP
The current sensor is not connected.
4
16
Over-Temp
Overheating is detected.
5
32
Not Used
Not used
6
64
Not Used
Not used
7
128
Not Used
Not used
8
256
Calibration failed
Calibration failure
9
512
PLL unlocked
PLL is unlocked.
*1
*1.
Unstable during in-rush current measurement (other measurement).
10
1024
PCR-COMM
Communication with the AC power supply cannot be
established.
11
2048
PCR-STAT
The AC power supply is not normal.
12
4096
SS
No steady state is found in the voltage fluctuation test.
13
8192
Not Used
Not used
14
16384
Not Used
Not used
15
32768
Not Used
Not used
Содержание KHA3000
Страница 20: ...20 KHA3000...
Страница 35: ...Overview This chapter provides an overview of this product and explains the options for the product 1...
Страница 42: ...42 KHA3000 Options...
Страница 49: ...Wiring Measurement Circuits This chapter explains the wiring of measurement circuits and the control circuit 3...
Страница 70: ...70 KHA3000 Connecting Communication Cables...
Страница 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Страница 91: ...Setting the Harmonic Test This chapter explains how to set the test conditions for the harmonic current test 5...
Страница 120: ...120 KHA3000 Controlling the AC Power Supply...
Страница 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Страница 247: ...Other Measurements This chapter explains basic measurement FFT analyzer and in rush current measurement 11...
Страница 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Страница 352: ...352 KHA3000 Cleaning and Checking...
Страница 353: ...Specifications This chapter explains the specifications of this product 15...
Страница 390: ...390 KHA3000...
Страница 391: ......