324
KHA3000
Querying Measurement Data
FETC[<channel#>][SCALar]:<meas_item>
READ[<channel#>][SCALar]:<meas_item>
MEASure[<channel#>][SCALar]:<meas_item>
FETC
Queries measurement data without starting measurement operation.
If FETC? is sent without measurement, an SCPI error (-230,
“
Data corrupt or stale
”
) occurs
because no measurement data is available. Always use this command with the INIT
command.
If <meas_item> is not specified, the value set by previous FETC?, READ? or
MEAS:<meas_item>? is used.
Command
FETCh[<channel#>][:SCALar][:<meas_item>]?
Response
Returns measurement data in the <NR3> format in response to FETC?.
READ
Queries measurement data after starting new measurement operation.
READ? works the same as the INIT command issued in combination with the FETC? query.
If <meas_item> is not specified, the value set by previous FETC?, READ? or
MEAS:<meas_item>? is used.
Command
READ[<channel#>][:SCALar][:<meas_item>]?
Response
Returns measurement data in the <NR3> format in response to READ?.
MEAS
Aborts the current measurement operation and starts the measurement operation anew, and
then queries measurement data.
MEAS:<meas_item>? works the same way as the ABORT command combined with the
READ? query.
Command
MEASure[<channel#>][:SCALar]:<meas_item>?
Response
Returns measurement data in the <NR3> format in response to MEAS:<meas_item>?.
•
If measurement has not been completed yet immediately after transmission of the INIT
command or the READ:<meas_item>? or MEAS:<meas_item>? query, the response data
FETC:<meas_item>?, READ:<meas_item>? or MEAS:<meas_item>? query is generated
after the completion of measurement.
Содержание KHA3000
Страница 20: ...20 KHA3000...
Страница 35: ...Overview This chapter provides an overview of this product and explains the options for the product 1...
Страница 42: ...42 KHA3000 Options...
Страница 49: ...Wiring Measurement Circuits This chapter explains the wiring of measurement circuits and the control circuit 3...
Страница 70: ...70 KHA3000 Connecting Communication Cables...
Страница 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Страница 91: ...Setting the Harmonic Test This chapter explains how to set the test conditions for the harmonic current test 5...
Страница 120: ...120 KHA3000 Controlling the AC Power Supply...
Страница 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Страница 247: ...Other Measurements This chapter explains basic measurement FFT analyzer and in rush current measurement 11...
Страница 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Страница 352: ...352 KHA3000 Cleaning and Checking...
Страница 353: ...Specifications This chapter explains the specifications of this product 15...
Страница 390: ...390 KHA3000...
Страница 391: ......