200
KHA3000
61000-3-11
●
Standard values of test impedance
Use these values when the current capacity of a power supply to be connected is 100 A or
more.
0.25
+ jn0.25
is for a single-phase EUT and 0.15
+ jn0.15
is for a three-phase
EUT.
This product cannot set the standard values of the above impedance. A selected or input
value in the menu is used for internal calculation.
No.
Menu item
Description
Default
1
Judge Factor
Select the condition item of Judge Factor. The items Pst and Plt cannot
be selected in manual measuring method.
Select the item with the small knob and press the ENTER key. Every
time the ENTER key is pressed, the check mark turns on or off. Multiple
checkboxes can be selected.
When multiple items are selected, the limit values of those items
become the conditions of the judgment.
Pst,
Plt,
dc,
dmax,
d(t)>3.3%
2
Return
Terminates the specification of Options of Judgment.
–
3
Z test
When the reference impedance in the 61000-3-3 standard is not used,
specify Test Impedance. Select the standard value 0.25
+ jn0.25
(single-phase) or 0.15
+ jn0.15
(three-phase) or
“
Specified.
”
0.25
jn0.25
4
R
A
test (
)
When
“
Specified
”
is selected in the F1 menu, input the resistance of
Test Impedance for each phase.
The input range is 0.00
to 1.00
.
Enter the value with the numeric keypad and press the ENTER key, or
use the large knob to increase or decrease the value.
0.24
5
X
A
test (
)
When
“
Specified
”
is selected in the F1 menu, input the reactance at the
fundamental frequency of Test Impedance for each phase.
The input range is 0.00
to 1.00
.
Enter the value with the numeric keypad and press the ENTER key, or
use the large knob to increase or decrease the value.
0.15
6
R
N
test (
)
When
“
Specified
”
is selected in the F1 menu, input the resistance of the
neutral line of Test Impedance.
The input range is 0.00
to 1.00
.
Enter the value with the numeric keypad and press the ENTER key, or
use the large knob to increase or decrease the value.
0.16
7
X
N
test (
)
When
“
Specified
”
is selected in the F1 menu, input the neutral line
reactance at the fundamental frequency of Test Impedance.
The input range is 0.00
to 1.00
.
Enter the value with the numeric keypad and press the ENTER key, or
use the large knob to increase or decrease the value.
0.10
8
Return
Terminates the specification of Test Impedance.
–
3
Input 1-char
Refer to Menu 3/4 of 61000-3-3 Standard.
–
4
Delete 1-char
–
5
Change char type
–
6
Return
–
7
Return
–
8
Std. Name Select
Standard
9
Input 1-char
–
10
Delete 1-char
–
11
Return
–
12
Return
–
9
Prev/Next
Page 3/4
Switches to another menu page.
–
F key: Function key. You can also use the ESC key to return.
Содержание KHA3000
Страница 20: ...20 KHA3000...
Страница 35: ...Overview This chapter provides an overview of this product and explains the options for the product 1...
Страница 42: ...42 KHA3000 Options...
Страница 49: ...Wiring Measurement Circuits This chapter explains the wiring of measurement circuits and the control circuit 3...
Страница 70: ...70 KHA3000 Connecting Communication Cables...
Страница 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Страница 91: ...Setting the Harmonic Test This chapter explains how to set the test conditions for the harmonic current test 5...
Страница 120: ...120 KHA3000 Controlling the AC Power Supply...
Страница 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Страница 247: ...Other Measurements This chapter explains basic measurement FFT analyzer and in rush current measurement 11...
Страница 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Страница 352: ...352 KHA3000 Cleaning and Checking...
Страница 353: ...Specifications This chapter explains the specifications of this product 15...
Страница 390: ...390 KHA3000...
Страница 391: ......