140
KHA3000
Estimating the Rsce (61000-3-12)
Searching for the Minimum Rsce Value with the Measured
Monitor Values Displayed (61000-3-12 2004 only)
Measured monitor values indicate the current status of the EUT. It is not a function to observe
all the operation cycles of the EUT.
Measured monitor value
●
Table format of measured monitor
The measured monitor table is formatted to always display L1, L2 and L3, irrespective of
equipment settings. L1, L2 and L3 correspond to LOAD terminals L1, L2 and L3 on the
rear of this product.
●
Sequ, Ssc and Z
The Sequ, Ssc and Z values are calculated for each phase of L1, L2 and L3 and
displayed. The formulas for computation are shown in the table above.
CH
Factor
Rsce
Sequ (VA)
Ssc (W)
Z (Ohm)
THD (%)
PWHD (%)
Phase of
measurement
circuit
L1
L2
L3
Harmonic
orders 2nd
to 40th,
THD,
PWHD
Short-
circuit ratio
Rated
apparent
power of
equipment
Short-
circuit
power
System
impedance
Total
harmonic
distortion
Partial
weighted
harmonic
distortion
–
Calculated
value
Calculated
value
Calculated
value
Calculated
value
Measured
value
Measured
value
Calculating
Formula
Rated apparent power (Sequ) of single-phase equipment = Rated voltage (phase voltage)
(Up)
×
Rated current (lequ)
Rated apparent power (Sequ) of three-phase/three-wire equipment = Rated voltage (line
voltage) (Ui)
×
Rated current (lequ)
Rated apparent power (Sequ) of balanced three-phase equipment =
×
Rated voltage
(line voltage) (Ui)
×
Rated current (lequ)
Rated apparent power (Sequ) of unbalanced three-phase equipment = 3
×
Rated voltage
(phase voltage) (Up)
×
Rated current (lequ)
Short-circuit power (Ssc) = Short-circuit ratio (Rsce)
×
(3
×
Rated apparent power (Sequ)
of single-phase equipment)
Short-circuit power (Ssc) = Short-circuit ratio (Rsce)
×
(2
×
Rated apparent power (Sequ)
of three-phase/three-wire equipment)
Short-circuit power (Ssc) = Short-circuit ratio (Rsce)
×
Rated apparent power (Sequ) of all
three-phase equipment
System impedance (Z) = (Nominal system voltage (Unom))
2
/Short-circuit power (Ssc)
Measured
monitor value
3
Содержание KHA3000
Страница 20: ...20 KHA3000...
Страница 35: ...Overview This chapter provides an overview of this product and explains the options for the product 1...
Страница 42: ...42 KHA3000 Options...
Страница 49: ...Wiring Measurement Circuits This chapter explains the wiring of measurement circuits and the control circuit 3...
Страница 70: ...70 KHA3000 Connecting Communication Cables...
Страница 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Страница 91: ...Setting the Harmonic Test This chapter explains how to set the test conditions for the harmonic current test 5...
Страница 120: ...120 KHA3000 Controlling the AC Power Supply...
Страница 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Страница 247: ...Other Measurements This chapter explains basic measurement FFT analyzer and in rush current measurement 11...
Страница 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Страница 352: ...352 KHA3000 Cleaning and Checking...
Страница 353: ...Specifications This chapter explains the specifications of this product 15...
Страница 390: ...390 KHA3000...
Страница 391: ......