112
KHA3000
61000-3-12
Setting with measured monitor values displayed (61000-3-12
2004 only)
The monitor values of real-time measurement are displayed in the upper part of the setting
item display. The measured monitor values indicate the current status of the EUT. This
function does not observe all operation cycles of the EUT.
6
Rated Current lequ (A)
Enter the rated current value of the EUT. For an unbalanced three-phase
device, enter the value of the maximum current that flows in one of the
three phases. This item is used for calculating the Sequ value.
The input range is 0.0 A to 75.0 A.
Enter the value with the numeric keypad and press the ENTER key, or
use the large knob to increase or decrease the value.
20.0 A
7
Nominal Sys Volt Unom
(
V
)
Enter a nominal system voltage. This item is used for calculating the
Ssc value.
The input range is 100 V to 600 V.
Enter the value with the numeric keypad and press the ENTER key, or
use the large knob to increase or decrease the value.
230 V
8
Nominal Frequency
Select the rated power frequency of the EUT.
50 Hz
9
Return
Terminates the specification of the rated voltage and frequency.
–
10
To Meas Setting Page
Switches to page 4/4.
–
11
Prev/Next Page 1/4
Switches to another menu page.
–
F key: Function key. You can also use the ESC key to return.
If you want to use the values (specific values) specified by the manufacturer in place of the
actually measured values for the rated current lequ, fundamental current (Iref), and standard
fundamental current (I1) of the EUT, select "Measured" to execute the test. Then, check that
the specified value is within ±10 % of the measured value, and execute the test using
specific values.
Write the test results using measured values in the Memo column of the report for tests
using specific values, or prepare each report separately and use them together.
Testing is possible even when the difference between measured values and specific values
exceeds ±10 %, but it is important to note that the limit value of the obtained test results does
not meet the standard.
Measured
monitor value
Содержание KHA3000
Страница 20: ...20 KHA3000...
Страница 35: ...Overview This chapter provides an overview of this product and explains the options for the product 1...
Страница 42: ...42 KHA3000 Options...
Страница 49: ...Wiring Measurement Circuits This chapter explains the wiring of measurement circuits and the control circuit 3...
Страница 70: ...70 KHA3000 Connecting Communication Cables...
Страница 71: ...System Setup This chapter explains how to turn on the power and how to set up the system 4...
Страница 91: ...Setting the Harmonic Test This chapter explains how to set the test conditions for the harmonic current test 5...
Страница 120: ...120 KHA3000 Controlling the AC Power Supply...
Страница 182: ...182 KHA3000 Checking the AC Power Supply and Repeatability...
Страница 247: ...Other Measurements This chapter explains basic measurement FFT analyzer and in rush current measurement 11...
Страница 349: ...Maintenance This chapter explains how to clean and inspect this product 14...
Страница 352: ...352 KHA3000 Cleaning and Checking...
Страница 353: ...Specifications This chapter explains the specifications of this product 15...
Страница 390: ...390 KHA3000...
Страница 391: ......