Execution error list
An error in the [-299, -200] range indicates that an error was detected by the
execution control block of the measuring instrument. If an error in this class occurs,
the execution error bit (bit 4) of the event status register is set.
Table 8-12 Execution error list
Query error list
An error in the [-499, -400] range indicates that the output queue control of the
measuring instrument detected a problem in the message exchange protocol
explained in IEEE 488.2 Section 6. If an error in this class occurs, the query error bit
(bit 2) of the event status register is set.
Table 8-13 Query error list
Error code
Explanation of error message
-200 Execution error (generic)
A generic error for this PAT
-203 Command protected
Password protected program or query command cannot be executed.
-210 Trigger error
A trigger error.
-211 Trigger ignored
A trigger was received but discarded.
-213 Init ignored
Measurement initiate operation was ignored because measurement is
in progress.
-214 Trigger deadlock
Dead lock occurred because a query was received before the
software trigger.
-220 Parameter error
A program data element related error occurred.
-221 Settings conflict
Received a command that cannot be executed in the current condition
of the PAT.
-222 Data out of range
Parameter value was outside the legal range.
-223 Too much data
Too many parameters were received than the requirements.
-224 Illegal parameter value
Received an invalid parameter data.
-230 Data corrupt or stale
Received a data query before the measurement was completed.
-241 Hardware missing
Cannot be executed because the optional hardware is not installed.
Error code
Explanation of error message
-400 Query error (generic)
A generic query error of this product that is used when other type of
error do not apply.
-410 Query INTERRUPTED
An INTERRUPTED query error as defined by IEEE 488.2 (6.3.2.3)
occurred.
-420 Query UNTERMINATED
An UNTERMINATED query error as defined by IEEE 488.2 (6.3.2.2)
occurred.
-430 Query DEADLOCKED
A DEADLOCKED query error as defined by IEEE 488.2 (6.3.1.7)
occurred.
-440
Query UNTERMINATED
after indefinite response
Another query is specified after a query that generates an undefined
response in a same program message.
(Example: SYST:ERR? is received after a *IDN? query and a
semicolon separator.)
Содержание KHA1000
Страница 1: ...Part No Z1 003 482 I0043901 Oct 2021 Operation Manual Harmonic Flicker Analyzer KHA1000 ...
Страница 8: ...vi KHA1000 ...
Страница 20: ...xviii KHA1000 Rear panel 1 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Страница 22: ...xx KHA1000 ...
Страница 23: ...Overview This chapter provides an overview of the product and explains its features 1 ...
Страница 30: ...1 8 KHA1000 ...
Страница 68: ...3 20 KHA1000 ...
Страница 69: ...Basic Operation This chapter explains basic operation methods 4 ...
Страница 156: ...5 54 KHA1000 ...
Страница 183: ...Other Measurements This chapter explains measurements for other than harmonic current and voltage fluctuation tests 7 ...
Страница 193: ...Remote Control This chapter explains how to connect the remote interface and how to control it using SCPI commands 8 ...
Страница 260: ...9 6 KHA1000 ...
Страница 261: ...Specifications This chapter explains the specifications of this product 10 ...
Страница 272: ...I 4 KHA1000 ...
Страница 273: ......