4-7
Remote Control
Query
The device settings or status can be queried.
To make a query, add a question mark at the end of the program header section.
If a query has parameters, enter a space after the question mark followed by the
parameters.
VOLTage:PROT? MIN
■
Response
A response returned as an answer to a query. It is a message that is always sent from
the device to the computer. The status of the device or measured values is
transmitted to the computer.
•
When transmitting two queries in separate lines, read the response to the first
query before transmitting the second line. When two query lines are sent at once,
an incomplete response may be received for the first query followed by a
complete response for the second query.
String termination
All commands must be terminated using a valid terminator.
The available terminators are <new line> (ASCII 0x0A) and EOI (end-or-identify).
Either one can be used as a terminator.
EOI is not available on the RS232C. Be sure to use <new line>.
When a command string is terminated, the path is reset to the root level.
•
CR (ASCII 0x0D) is not a terminator.
Common commands
The IEEE-488.2 and SCPI standards contain a set of common commands for reset,
self-test, and other functions. These common commands always start with an
asterisk. The commands may have one or multiple parameters.
Содержание KFM2151
Страница 1: ...OPERATION MANUAL FC Scanner Part No Z1 003 382 IB010911 May 2006 KFM2151...
Страница 4: ...ii KFM2151...
Страница 14: ...xii KFM2151 Contents...
Страница 19: ...General Description This chapter gives a general description and introduces the features of the KFM2151...
Страница 23: ...Installation and Preparation This chapter describes the procedures from unpacking to installation...
Страница 35: ...Operation This chapter mainly describes various setup procedures...
Страница 54: ...3 20 KFM2151 Operation...
Страница 81: ...KFM2151 4 27 Remote Control Fig 4 3 Status register SCPI Standard 1999 0 Volume1 fig 9 1...
Страница 100: ...4 46 KFM2151 Remote Control...
Страница 101: ...Maintenance This chapter covers maintenance of the KFM2151 including cleaning inspecting calibrating and troubleshooting...
Страница 105: ...Specifications This chapter contains the electrical and mechanical specifications of the KFM2151...
Страница 110: ...I 2 KFM2151 Index String 4 8 T terminator 4 7 U unit number setting of 3 13 V voltage protection 3 5...
Страница 111: ......
Страница 112: ...FC SCANNER KFM2151 OPERATION MANUAL...