KDS6-0.2TR
Remote Control 35
Fig. 4-3
Maps of status byte and service request enable registers
*CLS
Clears error registers and status byte registers excluding the RQS bit (bit 6). How-
ever, when a factor such as OCP or the like is ongoing, the relevant bit is set imme-
diately after clearing.
The RQS bit (bit 6) of a status byte register is cleared when serial polling is exe-
cuted or the power is turned off and on again.
Program message
•
Syntax
Command message:
*CLS
SRQ has
occurred
*STB
status byte register
Bit
7
6
5
4
3
2
1
0
Signal name
PON
RQS
ESB
MAV
ERR
–
–
OCP
Bit weight
128
64
32
16
8
4
2
1
*SRE
service request enable register
Bit
7
6
5
4
3
2
1
0
Signal name
–
–
–
–
ERR
–
–
OCP
Bit weight
128
64
32
16
8
4
2
1
PON: Setting of POWER ON SRQ (not in use)
ERR: ERROR register status summary
RQS: Request service
Not in use
ESB: Event status summary (not in use)
Not in use
MAV: Message ready (not in use)
OCP: OCP status summary
+
x
x
ERR: The bit becomes “1” if a command error or data section error occurs.
OCP: The bit becomes “1” if OCP is activated in any of Channels 1 to 3.
RQS: This bit becomes “1” when the bit registered in the service request enable
register becomes “1.”
Cleared when serial polling is performed.
Содержание KDS6-0.2TR
Страница 1: ...OPERATION MANUAL Precision DC Source Part No Z1 003 132 I0040701 May 2021 KDS6 0 2TR ...
Страница 4: ...ii KDS6 0 2TR ...
Страница 8: ...4 Safety Precautions KDS6 0 2TR ...
Страница 12: ...8 Introduction KDS6 0 2TR ...
Страница 22: ...18 Names and Functions of Parts KDS6 0 2TR ...
Страница 30: ...26 Basic Operations and Connections KDS6 0 2TR ...
Страница 52: ...48 Maintenance and Calibration KDS6 0 2TR ...
Страница 58: ...54 Index KDS6 0 2TR ...
Страница 59: ......