22
Basic Operations
KPM1000
Displayed contents
Selecting the display items for displays A, B, C, and D
Each time that you press DSPL A, DSPL B, DSPL C, or DSPL D, the measurement unit
changes to the next measurement unit in the following orders. Each time that you change the
measurement unit, the measurement item changes as well.
The orders that the measurement units change in are shown below. The KPM1000 combines
symbols and units to display units according to the display item.
Over-range indicators
If the value exceeds the specified value, the item shown on display A, B, C, or D changes.
V
DSPL A and DSPL C
DSPL B and DSPL D
A
W
Vpk
Apk
PF
Hz
[C]V
[C]A
[d]
V
A
W
Ah
[P]Wh
[n]Wh
Wh
VA
var
h
[ ]: Symbol display
Factory default settings
Display A: V
Display B: VA
Display C: A
Display D: W
Display item; [symbol display]; (unit)
Over-range indicator
Voltage (V)
If only a peak over-range occurs: P.oVEr. If only an RMS
over-range occurs: oVEr.
If a peak and RMS over-range occur: oVEr.
Current (A)
If only a peak over-range occurs: P.oVEr. If only an RMS
over-range occurs: oVEr.
If a peak and RMS over-range occur: oVEr.
Active power (W)
If only a voltage over-range occurs: oVEr.V. If only a current
over-range occurs: oVEr.i.
If a voltage and current over-range occur: oVEr.
Voltage peak (Vpk)
If only a peak over-range occurs: P.oVEr. If only an RMS
over-range occurs: Normal display.
If a peak and RMS over-range occur: oVEr.
Current peak (Apk)
If only a peak over-range occurs: P.oVEr. If only an RMS
over-range occurs: Normal display.
If a peak and RMS over-range occur: oVEr.
Power factor (PF)
If only a voltage over-range occurs: oVEr.V. If only a current
over-range occurs: oVEr.i.
Frequency (Hz)
If the frequency exceeds the upper measurement limit:
oVEr. If the frequency falls below the lower measurement
limit: undEr.
Apparent power (VA)
If only a voltage over-range occurs: oVEr.V. If only a current
over-range occurs: oVEr.i.
If a voltage and current over-range occur: oVEr.
Reactive power (var)
If only a voltage over-range occurs: oVEr.V. If only a current
over-range occurs: oVEr.i.
If a voltage and current over-range occur: oVEr.
Integrated current—sum—(Ah)
If the current display shows oVEr or P.oVEr during
integration, “Ah” blinks.
If the integrated value overflows: OVEr.F.
Integrated power—sum—(Wh)
If the voltage or current display shows oVEr or P.oVEr
during integration, “Wh” blinks.
If the integrated value overflows: OVEr.F.
Integrated power—positive—[P] (Wh) If the voltage or current display shows oVEr or P.oVEr
during integration, “Wh” blinks.
If the integrated value overflows: OVEr.F.
Integrated power—negative—[n]
(Wh)
If the voltage or current display shows oVEr or P.oVEr
during integration, “Wh” blinks.
If the integrated value overflows: OVEr.F.