DSI1020/ DSI3020
43
Installing the Test System
Turning the power on
Turn on the test system devices in the following order.
1
Turn on this product.
2
Turn on the POWER switches on the AC power supplies.
If a single-phase three-wire, three-phase three-wire, or three-phase four-wire output
test circuit has been constructed using several AC power supplies, turn on the phase
units in the order described on the next page. For details, see the setup guide of the
single-phase three-wire output or three-phase output driver that you are using.
Output from this product
Relays are inserted in the input circuit of this product as shown below. These relays switch to
on (dip simulator circuit) only when the dip simulator is enabled from the SD009-PCR-LE
Quick Immunity Sequencer 2 application software. For all other cases, the relays are set to
Bypass, which means that the AC power supply output is connected directly to the EUT. You
can perform tests without changing the wiring even when the dip simulator is not needed.
OUTPUT
INPUT
OUTPUT
PCR-LE
U
L
N
G
AUX(SLOT3)
SEQ TRIG OUT
On
Bypass
On
Bypass
On
Bypass
V
L
N
G
V
L
N
TRIG IN
AUX
TRIG OUT
PHASE IN
G
EUT
U
V
W
N
G
DSI3020
U
V
W
G
U
V
W
G
N
N
N
N
Connect to
PCR-LE and
DSI3020
PC
(SD009-PCR-LE)
Содержание DSI1020
Страница 62: ...62 DSI1020 DSI3020 This page has been intentionally left blank...