14
DSI1020/ DSI3020
Installing the Test System
Model DSI1020 rear panel
VOLTAGE
MONITOR
CURRENT
MONITOR
TRIG
OUT
TRIG
IN
RS232C
PHASE IN
AUX
ADDRESS
DO NOT REMOVE COVERS. NO OPERATOR SERVICEABLE PARTS INSIDE.
REFER SERVICING TO QUALIFIED SERVICE PERSONNEL.
SECURELY TURN OFF THE CIRCUIT BREAKER ON THE SWITCHBOARD BEFORE HANDLING THE POWER CORDS.
TO AVOID AN ELECTRIC SHOCK, THE PROTECTIVE CONDUCTOR TERMINAL MUST BE CONNECTED
TO AN ELECTRICAL GROUND.
WARNING
DSI1020
WEIGHT APPROX 110kg
AC300V 20A MAX
AC300V 20A
100-240V
50 60Hz 65VA MAX
AUX
1
AC INPUT
11
OUTPUT
12
INPUT
Base hold
angle
13
Protective conductor
terminal
10
VOLTAGE MONITOR
CURRENT MONITOR
TRIG OUT
6
7
8
TRIG IN
5
Slot for installing a
communication interface
(factory option)
9
PHASE IN
2
ADDRESS
3
RS232C
4
AC300V 20A MAX
AC300V 20A
100-240V
500 60Hz 65VA MAX
VV
14
Name
Function
AUX
An exclusive input terminal for receiving overheat protection signals (PCR CON-
TROL) from a LIN series line impedance network in a total test system.
PHASE IN
A control cable terminal for receiving sync clock signals from an AC power supply.
ADDRESS
Switches for selecting the communication interface (RS-232C, GPIB, or USB) and
setting the GPIB address.
RS232C
A connector for the RS-232C cable.
TRIG IN
An exclusive input terminal for receiving signals from AC power supplies to transmit
trigger signals during voltage variation tests.
TRIG OUT
An output terminal for transmitting trigger signals for starting and stopping voltage
dips and short interruptions and starting and stopping voltage variation.
VOLTAGE MONITOR
A signal output terminal for monitoring output voltage waveforms.
CURRENT MONITOR
A signal output terminal for monitoring output current waveforms.
—
A slot for installing a factory-option communication interface.
1
2
3
4
5
6
7
8
9